User Interface
35
Waltron User Manual 101-046-B.3
9096 Degassed Cation Conductivity Analyzer
Select Analog output 3
to bring up the
configuration for the
Degassed Cation
Conductivity signal
output.
Pressing the yellow box
to the right of Analog
Signal will show a list of
available output signal
types (4-20mA, 0-
20mA, 0-5V, etc).
Pressing the Scale Start
1 box will allow the user
to input a value that will
match the low range of
the output (4mA).
Pressing the Scale End
1 box will allow the user
to input a value that will
match the high range of
the output (20mA).
Pressing OK will return
to the previous menu
and save any changes
made to the
configuration.
Pressing Cancel will
return the previous
menu without saving
any changes that were
made.