Operating the analyzer
35
Waltron User Manual 101-045-E.1
9031CX Sodium Analyzer
3.5.7.1
RESULTS DATALOG
Results of your concentration readings by time and date are recorded in the Results Datalog. Data is
stored for up to 30 days at a sampling rate of 1 sample every 10 minutes. You may scroll up or down to
view results for the current day selected and select which day to view by selecting the day from the drop
down menu box in the upper right corner. Dates are displayed in YY.MM.DD format. Results are
recorded to the nearest hundredth of a ppb.
3.5.7.2
ALARMS DATALOG
The last 10 alarms are saved under the Alarms Datalog with red text. The user may press on the red text
to turn it green to indicate the alarm has been addressed. A description of all the alarms are given in
Section 3.2