Operating the analyzer
33
Waltron User Manual 101-045-E.1
9031CX Sodium Analyzer
3.5.6
CHECKS WINDOW
The “CHECKS” function allows you to perform maintenance diagnostics on the three way solenoid
valves, the three alarm relays, and the 4-20mA output.
To access the Checks Window:
From the Main Menu, press the “DISPLAY” button. Then press the “CHECKS” button. The buttons for
the valves and the relays will toggle on and off. The text within the button will turn red when the
valve/relay is open. To simulate the 4-20mA output press on the button and then enter a value as a
percentage into the white box using the keypad.