Operating the analyzer
32
Waltron User Manual 101-045-E.1
9031CX Sodium Analyzer
3.5.5
DISPLAY WINDOW
The display window will give you access to useful tools for troubleshooting. The display window will
show you the current analyzer reading graphically in mV as well as other useful information. The
analyzer can graph real time the last 10 minutes of the sensor reading in mV. Each vertical line represents
two minutes. The Y axis represents the sensor output in mV.
To access the Display Window:
Sensor:
Potential reading
Temperature:
Temperature reading of the sample at the flow cell
Offset:
Offset from the strength of the last calibration
Strength:
Percentage of mV1 reading vs mV2 (also referred to as slope)
Flow input:
Displays flow switch output (default ok if no flow switch is installed)
Electrode:
Displays user inputted vale for electrode identification (useful for recording date of electrode
replacement, etc)