Lid Lifts X*
The lid was lifted X times during the cycle.
Tank fluid level alert*
A problem was detected with the
cleaning fluid. This would normally be an
issue with the fluid level or excessive fluid
contamination.
Piezo fault alert*
The Piezo (Transducer) has a fault. The
unit should IMMEDIATELY be taken out of
service and returned for repair.
Electronics overheat alert*
The internal electronics have overheated.
This could be due to extended periods of
use or environmental conditions. If this
message regularly appears the unit
should be taken out of service and
returned for repair.
Fluid exceeded limit
temperature*
The unit was unable to control the
temperature and it rose above the preset
limit. The cycle therefore failed, as
proteins coagulate at temperatures in
excess of 45
o
C, which can lead to
inadequate cleaning of instruments.
Fluid temperature was
regulated
The fluid temperature reached the preset
limit and the power was reduced to
control this.
The supply was interrupted*
The mains supply was interrupted for
more than 7 seconds. If the power
interruption is less than 7 seconds the
unit continues the cycle.
Cycle FAILED
The cycle failed. All items above marked
* will cause a cycle to fail.
Cycle PASSED
All machine parameters were correct for
correct cleaning and the cycle passed.