Wahl DSTCAL Software User Manual 

 

 

13   

 
 

5.  SCREEN DEFINITIONS - DSX Models: 

Additionally to the previous standard definitions the DSX model software contains the 
following, reference Fig. 2: 
 
Module Configuration Information 
 
Module settings for S/N xxxxxxxx: 

4 mA Temp. (

C): 0.0

: Shows the equivalent current setting for the 4-mA output in 

F or 

C.  Temperature scale is changed under the “Settings” menu.  The 4-mA value must be 

a minimum of 10

C (18

F) below the 20-mA setting. 

 

20 mA Temp. (

C): 100.0

: Shows the current setting for the 20-mA output in 

F or 

C.  

Temperature scale is changed under the “Settings” menu.  The 20-mA value must be a 
minimum of 10

C (18

F) above the 4-mA setting. 

 

Burnout Setting: 22.0 Upscale (22.0 mA)

: User selectable setting for fault indication to 

drive the current to a low or high setting in the event a sensor gets disconnected, or 
opens.  The current values are selected to comply with the NAMUR recommendation 
NE43 (18.01.1994).  The two settings available from the drop down menu are: 
 

Upscale (22.0 mA)

 

– drives the current to 22.0 mA if fault is detected. 

Downscale

 

(3.5 mA) 

– drives the current to 3.5 mA if fault is detected. 

 

Calibration Information 

 
Module Calibration Data 

 
4-20 mA Cal. Slope:

 0.000244 - Displays slope correction for D/A converter. 

 

4-20 mA Cal. Offset:

 -0.00925 

– Displays offset correction for D/A converter. 

 
 

Select Calibration to Perform

: For DSX models, there is a third option on the “Select 

Calibration to Perform” drop down box that allows the user to perform calibration of the 
current loop output.  Select the “4-20 mA Calibration” and click the “Calibrate” button.  A 
calibration routine with step by step instructions will take you through the calibration 
process.   
 

4-20 mA Calibration

 

– Performs calibration of the electronics for the 4-20 mA current 

output.  This calibration requires a precision current meter with resolution of 1 microamp 
or less. 

Summary of Contents for DST400

Page 1: ...oftware User Manual Palmer Wahl Instruments Inc 234 Old Weaverville Road Asheville NC 28804 Toll Free 800 421 2853 Phone 828 658 3131 Fax 828 658 0728 www palmerwahl com 09 16 09 WD1037 Rev F Revised...

Page 2: ...els 13 6 CALIBRATION DSX Current Loop 15 6 1 Theory of Calibration 15 6 2 Current Loop Calibration 15 4 20 mA Calibration Window 15 7 DST600 PROBE ID 16 7 1 Theory of Operation 16 7 2 Definitions 16 1...

Page 3: ...Wahl DSTCAL Software User Manual 3...

Page 4: ...ion on the DSX models Association and enabling of probe ID s for DST600 series Models Setting of Low Battery Check rate Calibration of the meter s measurement electronics resistance input for RTD mode...

Page 5: ...B Interface Module 2 Connect the 4 pin cable assembly p n 13507 to the DST USB Interface Module port labeled DIGI STEM 3 With the Digi Stem unit open connect the opposite end of the 13507 cable to J3...

Page 6: ...the 4 mA and 20 mA values DSX models only SPY COM Serial Communication Debug Tool Not used during normal operation Help About Displays software version and contact information for Wahl Instruments Sel...

Page 7: ...rature from the measured resistance For the highest accuracy of measurements it should be set to match the R0 value of the probe being used It may be manually entered or may be automatically calculate...

Page 8: ...ettings are the calibration coefficients used in calculating temperature from the measured probe resistance These may be changed by the user via the Ohm Meter Calibration routine Restore Factory Setti...

Page 9: ...fficient To increase the total system accuracy the electronics may be programmed with this value so that the microprocessor can accurately calculate the actual temperature from the measured resistance...

Page 10: ...et Factory Calibration Window DSTCAL Pro only 14 Do you wish to calibrate the module with the values below Set Factory calibration pair to xxx xxxx x xxxx 15 Click Yes to continue or No to return to F...

Page 11: ...ter Calibration Probe system Calibration Window 1 Click on the Select Calibration to Perform drop down box and select Probe system Calibration 2 Click the Calibrate button 3 Placing probe in Ice Bath...

Page 12: ...d if the input resistance is not within approximately 4 ohms of the requested nominal setting One or both of the following error messages may occur when in the Meter Only Calibration and the calculate...

Page 13: ...cted or opens The current values are selected to comply with the NAMUR recommendation NE43 18 01 1994 The two settings available from the drop down menu are Upscale 22 0 mA drives the current to 22 0...

Page 14: ...Wahl DSTCAL Software User Manual 14 Fig 2...

Page 15: ...f value is out of acceptable range 2 Click Next to continue or Cancel to abort Enter Measured Current 20 mA Screen 3 Forces DSX unit to output 20 mA signal and prompts user to enter actual value measu...

Page 16: ...ed probe is being used To authorize a new probe the unit must be connected to the programming or calibration software where it is recommended the unit be recalibrated to ensure accuracy The Probe ID m...

Page 17: ...ollowing screen will appear Clicking OK will close this window To associate the new probe click the Capture Probe button Fig 3 The following screen will appear To associate the new probe clicks yes ot...

Page 18: ...temperature scale to be used when programming the 4 mA and 20 mA values DSX models only currently not available on 400 series Help About Displays software version and contact information for Wahl Ins...

Page 19: ...y checking Thermocouple Type Allows selection of J K E T or S thermocouples Changing of the thermocouple type requires the recalibration of the Cold Junction Compensation using the newly selected ther...

Page 20: ...to insert a 25 C equivalent signal via thermocouple wire of the selected type to the DST400 input connector This calibration must be performed after changing thermocouple type Existing Thermocouple Ca...

Page 21: ...le type wire The correction factor used by the meter is the mV equivalent of the temperature offset measured in accordance with the selected thermocouple type s mV to temperature table 9 2 Meter Calib...

Page 22: ...d to Set Factory Calibration window below Set Factory Calibration Window DSTCAL Pro only 13 Do you wish to calibrate the module with the values below Set Factory calibration pair to xxx xxxx x xxxx 14...

Page 23: ...Screen 5 Displays results of calibration 6 Click Set Calibration to continue Clicking Cancel will return to main screen Set Active Calibration Window 7 Do you wish to calibrate the module with the val...

Page 24: ...L USB 232 interface box and should be selected in the select COM port window 6 Plugging the cable back in will force the list to refresh again and it should now include the COM port for the interface...

Page 25: ...the DSTCAL program 4 Restart the computer 5 Open DSTCAL 6 Connect the DST USB Interface Module to the computer 7 Install the battery into the Digi Stem 8 Connect the DST USB Interface Module to the D...

Reviews: