DESCRIPTION OF THE INTERNAL WATER TANKS
The sterilizer is equipped with two independent water tanks, one for clean water and one for used water.
The tank on the right side is called "clean water tank" and has to be filled with distilled or demineralized
water required for the sterilization process. It is fitted with a minimum (0.6l) and a maximum (4 l) water lev-
el sensor. The tank can be filled manually through the tank hole on the top of the sterilizer or automatically
with an external water supply system (see § 4.5.1).
Use the quick connector behind the service door (left/blue) to drain the clean water tank for cleaning pur-
poses (see chapter on maintenance).
The tank on the left side is called "used water tank" and contains the used water collected at the end of
each sterilization cycle. It is fitted with a maximum water level sensor (3.5l).
Use the quick connector behind the service door (right/grey) to drain the used water tank (see chapter on
maintenance).
The used water tank can also be drained automatically by connecting a drain tube to the D4 fitting located
at the back of the sterilizer (see §4.5).
NOTE:
The water consumption per sterilization cycle varies depending on the type and weight of the
sterilization load. The capacity of the clean water tank is sufficient to run 7-11 sterilization
cycles.
Used water tank
Water level sensors
Quick connector for
manual draining of
clean water tank
Quick connector for ma-
nual draining of used
water tank
Water inlet for manual filling of
clean water tank
Fittings for automatic fill-
ing/draining of tanks
Clean water tank
Vent