37
Therapy Protocols
You can access, create, and manage therapy protocols from the Main Screen
prior to interrogation. This allows you to defi ne the mode, output current,
and parameter settings before you begin working with a patient.
To access therapy protocols:
1.
From the Main Screen, tap on the
SETTINGS
button in the
Navigation Bar at the bottom of the screen
2.
Tap on the
GUIDED MODE OPTIONS
button
3.
From the Guided Mode Options screen, tap on
THERAPY PROTOCOLS
button
To create a custom protocol:
1.
From the Therapy Protocols screen, select
CREATE PROTOCOL
2.
From the Create Protocol - Select a Starting Point screen, select
START FROM SCRATCH
or choose an existing protocol as base
3.
Add or delete steps (maximum of 7) and set the output currents for
each therapy mode. When you are fi nished, tap
NEXT
4.
Choose the custom persistent parameters that will be used in all
protocol steps. When you are fi nished, tap
NEXT
5.
Follow on-screen prompts to name the custom protocol.
When you are fi nished, tap
SAVE
6.
From the Therapy Protocols Screen, use the
BACK
arrow to
return to the Main Screen
Summary of Contents for SENTIVA M2000 Wand
Page 1: ...1 Technical Guide ...
Page 38: ...38 Tap to delete steps ...
Page 44: ...44 NOTE To view more details tap on the individual bars in the histogram ...
Page 54: ...54 ...