33
Scheduled Programming
The generator must already be in Guided Mode before Scheduled Programming
can be used. Use Scheduled Programming to automatically change parameter
steps according to a preset schedule. Scheduled programming may be useful to
patients who live far from their healthcare provider.
To enable Guided Programming:
1.
From the Summary Screen, tap on the
EDIT PARAMETERS
button, or tap on
PARAMETERS
in the Navigation Bar at the bottom of the screen
2.
On the Parameters Screen, tap on
SWITCH TO GUIDED MODE
button at
top of screen
3.
On the Switch to Guided Mode Screen, select either
STANDARD
or
CUSTOM
protocol and tap
NEXT
4.
Tap
APPLY CHANGES
to enable Guided Mode
To turn on Scheduled Programming:
1.
In Guided Mode View, tap the
SCHEDULED-APPLY
button
2.
On the Scheduled-Apply Screen, tap on
ENABLE NOW
at the center
of the screen
3.
On the next Scheduled-Apply screen, select the time interval between step
increases, then tap
GENERATE SCHEDULE
at the center of the screen.
Once a schedule has been generated you can follow the on-screen prompts
to make adjustments to the schedule
4.
Tap
APPLY CHANGES
to program your changes
NOTE:
At the end of each offi ce visit, the generator must be interrogated by
tapping the
INTERROGATE
button followed by tapping the
END SESSION
button to ensure creation and accuracy of the session reports.