5.3.4 Z-axis
The z-axis in lubricated through 4 nipples located on both sides of the axis..
Pull off the black plastic caps and lubricate the axes using the lubrication nipples which have
been exposed. Replace the plastic caps after finishing the procedure.
Figure 3: Access lubrication Z-axis
(1)
Plastic cap. Underneath the cap the nipples for lubrication are located.
There are two more nipples located at the opposite side of the axis.
At the end of the lubrication process please plug in the power and switch on again the
system at the main switch.
Please start the control computer and unlock the laser with the green power button again.
Then move all 3 axis by using the "VitroMark" software 2 times over the entire range of
motion to distribute the lubricant.