™ Chapter 3: Functioning
- 71 -
Once the point of best focus has been reached, the arrows will be replaced by a
green mark, indicating that the machine is already at the optimum distance, so you
should not move the machine any more.
Figure 46. Green mark indicating the position of best focus.
It is advisable, during this focusing process, that the patient, although he/she can
blink, should avoid doing so excessively, since blinking may confuse the program.
If at any time during the focusing process the program stops showing the directional
arrows, cancel the operation completely, and start the measurement again.