HD Analyzer
™ Chapter 2: Hardware installation
- 19 -
2 HD
A
NALYZER
™
H
ARDWARE
I
NSTALLATION AND
M
AINTENANCE
2.1
HD ANALYZER™ INSTALLATION
Installation of
the HD Analyzer™ equipment must be carried out by an authorized
installer.
The equipment is not designed to be portable. Once installed, moving it to another
location is discouraged. If it is necessary to move it to a nearby location, we
recommend that two people move it, lifting the instrument by its base, each person
on one side of the instrument.
Never seize the HD Analyzer
™ except from its base. The moving
part is very fragile and could damage the interior elements.
If transport to a distant location is necessary, you should contact your local
distributor or the manufacturer to carry out the packaging process correctly and the
subsequent installation at the new destination.
2.2 MAINTENANCE AND CALIBRATION
Calibration
It is recommended that you contact your distributor to perform a calibration once a
year. The user must not carry out any calibration work on this equipment, as this
must be carried out by specialized personnel.
The unit label contains the date of the next recommended calibration. In addition,
the program itself, in its setup, stores that date, so it will notify you 30 days before
that date, or when it detects that date has already passed.