BewellConnect - BW-TSX/BW-TS1 - User Manual - 11-2017v2
BewellConnect - BW-TSX/BW-TS1 - User Manual - 11-2017v2
28
29
RF emissions
CISPR 11
Group 1
The BW-TSX / TS1 use RF
energy only for its internal
function. Therefore, its
RF emissions are very low
and are not likely to cause
any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The BW-TSX / TS1 is suitable
for use in all establishments,
including domestic
establishments and those
directly connected to the
public low-voltage power
supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
GUIDANCE AND MANUFACTURE’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The BW-TSX / TS1 is intended for use in the electromagnetic environment specified below.
The customer or the user of BW-TSX / TS1 should assure that it is used in such an
environment.
IMMUNITY
TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNE
TIC ENVIRONMENT
- GUIDANCE
Electrostatic
discharge
(ESD)
IEC
61000-4-2
±6 kV contact
±8 kV air
±8 kV contact
±15 kV air
Floors should be
wood, concrete or
ceramic tile. If floor
are covered with
synthetic material,
the relative humidity
should be at least
30%.
Electrical fast
transient/burst
IEC
61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2kV for power
supply lines
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Surge
IEC
61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
±1 kV differential
mode
Mains power quality
should be that of a
typical commercial
or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on power
supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 5 sec
Mains power quality
should be that of a
typical commercial
or hospital
environment. If the
user of the BW-TSX
/ TS1 requires
continued operation
during power mains
interruptions, it is
recommended that
the BW-TSX / TS1
be powered from
an uninterruptible
power supply or a
battery.
Power
frequency
(50Hz/60Hz)
magnetic field
IEC
61000-4-8
3 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
GUIDANCE AND MANUFACTURE’S DECLARATION – ELECTROMAGNETIC IMMUNITY
The BW-TSX / TS1 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BW-TSX / TS1 should assure that it is used in such an
environment.
IMMUNITY
TEST
IEC 60601 TEST
LEVEL
COMPLIANCE
LEVEL
ELECTROMAGNETIC
ENVIRONMENT - GUIDANCE