7-2 Troubleshooting
-81-
Problem Cause Remedy
The balance may have been
affected by an external influence
such as wind and vibration
during calibration.
The balance may be affected by
an external influence such as
wind and vibration. Take
remedial actions or relocate the
balance by referring to the
section “Do not use the balance
in the following places” (P. v).
Then calibrate again.
Weight indication contains an
error even after calibrated.
The weight used for calibration
is slightly different in mass from
the weight used for checking.
Use the same weight during
calibration and checking.
The display does not move with
the M sign flashing.
(When the [Zero/Tare] key is
pressed, during a sampling in
parts counting mode, etc.)
The balance may be affected by
an external influence such as
wind and vibration.
The balance may be affected by
an external influence such as
wind and vibration. Take
remedial actions or relocate the
balance by referring to the
section “Do not use the balance
in the following places” (P. v).
The icon blinks when the balance
is operated on batteries.
The batteries are exhausted.
Replace the batteries.
The display is turned off when
the balance is operated on
batteries.
The display was turned off by the
auto power off function.
The auto power off function is
activated if the balance is left
unused with no measurement
taken for approximately five
minutes.
Disable the auto power off
function if it interferes with your
use of the balance.