background image

6-2  Connecting to External Devices via RS-232C Interface 

-65- 

参 考

 

You can set a tare range or adjust the zero-point from an external device by connecting a 

contact or a transistor switch between the pin for externally setting a tare range (Pin 9) to 

the pin for signal ground (Pin 5). In this case, allow at least 400 ms for connection (ON) 

time (Maximum voltage when the balance is turned OFF: 15 V, sink current when it is 

turned ON: 20 mA.) 

 

Sample connection with a PC 

Use the following examples as a guide to connect the balance to external devices using the cable. 

 

-  Sample connection with a PC/AT compatible machine 

 

TXD

RXD

GND

DCD

RTS

CTS

DSR

DTR

RXD

TXD

GND

3

2

5

1

7

8

6

4

2

3

5

パソコン側

(D-SUB9P)

はかり側

(D-SUB9P)

 

 

TXD

RXD

GND

DCD

RTS

CTS

DSR

DTR

RXD

TXD

GND

2

3

7

8

4

5

6

20

2

3

5

パソコン側

(D-SUB25P)

はかり側

(D-SUB9P)

 

 

PC 

Balance 

PC 

Balance 

Reference 

Summary of Contents for CJ-E series

Page 1: ...Electronic Balance CJ E Series Operation Manual Instructions To ensure safe and proper use of the balance please read this manual carefully After reading this manual store it in a safe place near the...

Page 2: ......

Page 3: ...rgency alerted for danger is high WARNING This symbol indicates improper handling that may cause death or severe damage including serious injury CAUTION This symbol indicates improper handling that ma...

Page 4: ...rvice Division Only use the specified power source Use of other types of power sources may result in heat generation fire or malfunction of the balance Only use the dedicated AC adapter Use of other t...

Page 5: ...dusty environment Do not lay the AC adapter cable on the surface of a passage Somebody may trip on the cable causing the balance to fall thereby causing injury and or damage to the balance Do not move...

Page 6: ...vent breakage or malfunction place the sample to be measured on the balance carefully Do not use volatile solvents for cleaning The body may be distorted To clean the unit of stains use a piece of dry...

Page 7: ...with 80 or lower relative humidity Location where it is subject to direct sunlight An internal temperature increase in the balance may lead to inaccurate measurement Unstable base or location where i...

Page 8: ...wer and to prevent deterioration unplug it CAUTION battery handling Never disassemble or modify the batteries Take care to ensure you insert batteries with the positive and negative poles correctly in...

Page 9: ...Safety Precautions vii Dispose of batteries in accordance with local regulations...

Page 10: ...setting units and minimum readability Chapter 6 Input Output to from External Devices This chapter describes printing to printers and how to input and output to from RS 232C devices in detail Chapter...

Page 11: ...Displaying the Sum of the Container and the Sample 15 2 5 Function Setting Basics 16 3 1 Weighing Weighing Machine 20 3 2 Counting Parts Count 21 Displaying average sample weight 22 3 3 Measuring Perc...

Page 12: ...by deviation values using the method of entering values 44 4 1 How to Calibrate 48 4 2 Testing the Balance 50 5 1 Using Two Expression Units by Switching Them 52 5 2 Minimum Readability Setting 53 5...

Page 13: ...ing 80 7 3 Initializing 83 7 4 Maintenance 84 For heavy dirt 84 How to take care of the balance 84 Appendix 1 Function Setting List 86 Appendix 2 Measurement Mode List 90 Appendix 3 Printing in Compli...

Page 14: ......

Page 15: ...assembling the main unit to turning the power switch on and off Before your first use of the balance be sure to read this chapter This chapter includes Checking Supplied Items Names and Functions of...

Page 16: ...n the box In the unlikely event of problems such as missing or broken items please contact the retailer from whom the balance was purchased or our Sales Office See the Appendix at the end of this manu...

Page 17: ...inclination of the unit by turning them right or left D SUB9 pin Use to connect the balance to an RS232C connector printer or other external device Cover Remove when using options Anti theft hole Con...

Page 18: ...ust from entering the unit Note that the waterproofing and dustproofing functions do not work in the following situations When the cable is connected to the D SUB9P When the buzzer option the full pac...

Page 19: ...s Press Give a light press of the key and then release the finger Press and hold Hold the key down and release the finger after the appropriate display is obtained Continuously press and hold Hold dow...

Page 20: ...that settings are being saved Blinks when adjusting the zero point and when waiting for tare range setting to stabilize Zero point Bar graph Shows gross weight using the rate to the weighing capacity...

Page 21: ...mode Upper Indicates that an ID number is being displayed or entered Indicates that actual water temperature unit C is being entered Upper Indicates that a specific gravity unit none is being displaye...

Page 22: ...sing a tool such as a coin Attach the pan base by aligning it to the projections on the main unit Do not overtighten the screws Fix it by turning the screws using a tool such as a coin 2 Attach the pa...

Page 23: ...ition and make adjustments little by little Making adjustments using the other adjuster located on the other side as a pair while observing the level helps to level the balance After the air bubble in...

Page 24: ......

Page 25: ...c measuring functions that are used daily This chapter includes Powering On Off the Balance and Checking Operation Weighing by Placing a Sample in a Container Tare Weighing an Added Sample Displaying...

Page 26: ...n the power is turned on is the measuring mode that was used before the power was turned off For example if the power is turned off in parts counting mode the balance is started up in parts counting m...

Page 27: ...container is displayed 2 Set the tare 0 Net Press the Zero Tare key The readout becomes zero and 0 and Net are displayed tare range setting When Net is not displayed even though the Zero Tare key is p...

Page 28: ...Only the added weight is measured 1 Place the sample Net The weight of the placed sample is displayed 2 Set the tare 0 Net Press the Zero Tare key The readout becomes zero tare range setting 3 Place t...

Page 29: ...mode refer to Weighing P 20 1 Place the container and then set the tare Place the container and press the Zero Tare key The tare range is set and the readout becomes zero 2 Place the sample The weigh...

Page 30: ...the balance to another mode If this happens press the Print key to cancel the setting and redo the setting from the beginning 2 Select a setting item By pressing the Function key select an item to be...

Page 31: ...the function setting refer to Section 7 3 Initializing P 83 Function setting is possible with the direction keys After switching to the function setting mode with step 1 use the direction arrow keys...

Page 32: ......

Page 33: ...ng percentage weighing unit converting gravimeter and animal weighing This chapter includes Weighing Weighing Machine Counting Parts Count Measuring Percentage Obtaining Weight Multiplied by Coefficie...

Page 34: ...use the following operation 1 Set to the function setting mode Refer to Section 2 5 Function Setting Basics P 16 Press and hold the Function key After is displayed release the finger is displayed 2 S...

Page 35: ...ease the finger 2 Select Parts Counting Press the Zero Tare key several times to select 3 Save the weighing mode Pcs Press the Set key The parts counting mode is set displaying Pcs 4 Start sampling Pr...

Page 36: ...sample weight When weighing in parts counting mode pressing the Function key can display average sample weight Pressing the Function key toggles the display between number of samples average sample w...

Page 37: ...y After is displayed release the finger 2 Select Percentage Weighing Press the Zero Tare key several times to select 3 Save the weighing mode Press the Set key The percent weighing mode is set display...

Page 38: ...d hold Press and hold the Function key After is displayed release the finger The previously saved reference sample weight blinks 5 Enter a reference weight Set the reference value with the following s...

Page 39: ...the saved reference weight Min Indication Range of Reference Weight 1 Lower weight limit Reference weight Lower weight limit 10 0 1 Lower weight limit 10 Reference weight Lower weight limit 100 0 01 L...

Page 40: ...Tare key several times to select 3 Save the weighing mode Press the Set key The unit converting mode is set displaying 4 Set to the coefficient setting mode Press and hold Press and hold the Function...

Page 41: ...ave the coefficient Press the Set key 7 Weigh The weight of the sample is multiplied by the coefficient and the result is displayed The increment of the minimum readability is automatically set to 1 2...

Page 42: ...weight of the sample in the air 4 Correcting errors due to the cage 5 Measuring the weight of the sample in water 6 Displaying a specific gravity value Preparing measurement equipment Prepare the fol...

Page 43: ...ainer water temperature Set the water temperature with the following steps Set the water temperature starting from a higher order digit 1 Press the Zero Tare key Zero blinks in the rightmost digit 2 S...

Page 44: ...n the air When the weight is saved is displayed in the lower left of the display Measuring a sample by placing it on the pan is also allowed 9 Set the liquid container Set the container with water und...

Page 45: ...cm3 Gravimeter measurement data output Measured specific gravity data is output to a printer as follows depending on settings Before measurement Irrespective of the setting made in the function setti...

Page 46: ...Various Measuring Methods 32 Set language for printing by the function setting language for printing 1 English 2 Japanese...

Page 47: ...ta output samples When water is selected English Japanese katakana Specific gravity of solid Sample weight Actual water temperature When other than water is selected English Japanese katakana Specific...

Page 48: ...played release the finger 2 Select Animal Weighing Press the Zero Tare key several times to select 3 Save the weighing mode g Press the Set key The animal weighing mode is set displaying 4 Set the res...

Page 49: ...If the animal moves too much the value may not be held The measurement unit used in animal weighing is g only In animal weighing because stability detection ranges are wide errors may occur compared...

Page 50: ...setting mode Refer to Section 2 5 Function Setting Basics P 16 Press and hold the Function key After is displayed release the finger 2 Select Addition Function Press the Function key several times to...

Page 51: ...display indicates 0 and place the next sample Cumulating weight is also possible as follows Press the Zero Tare key without unloading the balance and then place next samples 3 Display the cumulative...

Page 52: ...to 2 point bar graph Refer to Appendix 1 Function Setting List P 86 Bar graph display is possible only for two point setting Judgment criteria and limit value setting A limit value can be judged with...

Page 53: ...re judged 2 Upper and lower limits are set HI OK and LO are judged Judge by 1 Judge by absolute values 2 Judge by deviation values Buzzer for LO 0 The buzzer is not beeped for rank LO 1 The buzzer is...

Page 54: ...judge even when the balance is unstable 2 Judge only when the balance is stable 4 Set the judgment range to cover Press the Function key several times to select Press the Zero Tare key to select the j...

Page 55: ...of using actual samples Press and hold Press and hold the Set key After is displayed release the finger 2 Weigh the lower limit sample Place the sample serving as the lower limit and then press the Fu...

Page 56: ...2 Select a number by pressing the Zero Tare key Pressing the key toggles the number between 0 and 9 and decimal point 3 Pressing the Function key shifts the blinking digit to the left and the next lo...

Page 57: ...finger 2 Weigh the reference weight sample Place the sample serving as the reference value and then press the Function key 3 Weigh the lower limit sample Place the sample serving as the lower limit an...

Page 58: ...ues Press and hold Press and hold the Set key After is displayed release the finger 2 Enter the reference value Enter the reference value with the following steps 1 Press the Zero Tare key Zero blinks...

Page 59: ...3 8 Judging Above and Below Limit Function 45 5 Weigh Place a judgment target sample on the pan The OK LO or HI result judged from the upper and lower limit values is displayed with the indication...

Page 60: ......

Page 61: ...Chapter 4 Adjusting the Scale Calibrate the balance using weights This chapter includes How to Calibrate Testing the Balance...

Page 62: ...calibrate more accurately use a weight that is equivalent to the weighing capacity Please contact us if you wish inquire about or place an order for calibration weights 1 Set to the function setting m...

Page 63: ...djustment starts 8 Span adjustment finishes After span adjustment finishes is displayed The balance goes back to normal display Pressing any other key than the Function key cancels adjustment The disp...

Page 64: ...er is displayed release the finger 2 Set to the span test mode Press the Function key several times to select Press the Zero Tare key several times to select 3 Save the setting Press the Set key The b...

Page 65: ...g Set the functions of the balance This chapter includes Using Two Expression Units by Switching Them Minimum Readability Setting Saving Container Tare Weight Power Setting ID No Setting Improving the...

Page 66: ...ease the finger 2 Set unit A Press the Function key several times to select Select unit from the following numbers by pressing the Zero Tare key g kg ct carat etc When setting only the unit A press th...

Page 67: ...es to select Select from 1 through 5 by pressing the Zero Tare key 3 Save the setting Press the Set key The balance goes back to normal display To set the minimum readability of unit B select in step...

Page 68: ...ghing You should set the tare range regularly 1 Set to the function setting mode Refer to Section 2 5 Function Setting Basics P 16 Press and hold the Function key After is displayed release the finger...

Page 69: ...layed release the finger 2 Set the auto power off Press the Function key several times to select Select 1 by pressing the Zero Tare key 3 Save the setting Press the Set key The balance goes back to no...

Page 70: ...er is displayed release the finger 2 Set the auto backlight off Press the Function key several times to select Select 1 by pressing the Zero Tare key 3 Save the setting Press the Set key The balance g...

Page 71: ...you can use are as follows Space blank 0 to 9 A to F 1 Set to the function 2 setting mode Press the Function key while pressing the Zero Tare key Release the fingers when is displayed The function 2...

Page 72: ...Function Setting 58 5 Save the ID number Press the Set key Display changes to Pressing the Set key again returns the balance to normal display...

Page 73: ...and wind vibration influences Wind vibration influences Stability judgment Response speed Minimum readability setting 1 1 1 2 2 2 3 3 3 4 4 4 Small Big 5 5 In each of the functions if wind and vibrati...

Page 74: ...Function Setting 60 4 Save the setting Press the Set key The balance goes back to normal display...

Page 75: ...Balance data can be output to a printer and be input and output to from external devices via the RS 232C interface This chapter includes Outputting to a Printer Connecting to External Devices via RS...

Page 76: ...e operation manual of the printer Set the printing function printing control to balance control Make the baud rate and other communication settings compatible with the settings made in the balance The...

Page 77: ...efer to Section 2 5 Function Setting Basics P 16 Press and hold the Function key After is displayed release the finger 2 Press the Function key several times to select Select 1 by pressing the Zero Ta...

Page 78: ...established with the following specifications Disconnect the AC adapter of the balance before connecting external devices Connector pin numbers and functions The RS 232C connector equipped on the bala...

Page 79: ...ast 400 ms for connection ON time Maximum voltage when the balance is turned OFF 15 V sink current when it is turned ON 20 mA Sample connection with a PC Use the following examples as a guide to conne...

Page 80: ...nt with EIA RS 232C HIGH level data logic 0 5 to 15 V LOW level data logic 1 5 to 15 V Bit configuration Start bit One bit Data bits 8 7 bits 7 bit can be specified only for the extended 7 digit numer...

Page 81: ...ng mode Refer to Section 2 5 Function Setting Basics P 16 Press and hold the Function key After is displayed release the finger 2 Press the Function key several times to select Press the Zero Tare key...

Page 82: ...e one bit rather than two bits If you select Japanese katakana for printed language the data length is automatically set to eight bits Meanings of data P1 one character Indicates the polarity of data...

Page 83: ...H pound O T 4FH 54H troy ounce D W 44H 57H pennyweight G R 4BH 52H Grain Bottom right grain T L 54H 4CH tael Hong Kong T L 54H 4CH tael Singapore Malaysia Top right T L 54H 4CH tael Taiwan Middle righ...

Page 84: ...git numeric format 3000 1 g No data type specified Data stable 1 2 3 4 5 6 7 8 9 10 11 12 13 14 0 3 0 0 0 1 SP G SP S CR LF 800 05 mom Gross Data unstable 1 2 3 4 5 6 7 8 9 10 11 12 13 14 0 8 0 0 0 5...

Page 85: ...thin one second after it is received For a tare range command and a span adjustment test command a response is sent after the command is completely processed If the balance receives a tare range comma...

Page 86: ...zero point adjustment command C1 C2 Code C1 Code C2 Description Value Response T SP 54H 20H Tare range setting Zero point adjustment None A00 Successful completion E01 Command error E04 A tare range...

Page 87: ...e when the balance is stabilized again the output includes zero even if it is not reloaded O 6 4FH 36H Output once when the balance is stable Output continuously when unstable Output is stopped after...

Page 88: ...rror Mode 4 Display in unit B 2 Average unit weight Error Error Error Error 1 Mode 3 M3 can be specified only when the addition function is to be used If the addition function is not enabled it result...

Page 89: ...terminators For more information on A1 to A3 see the Response fields in the previous section A1 A2 A3 CR LF ACK NAK format Consists of one character without a terminator Successful completion ACK or...

Page 90: ......

Page 91: ...Chapter 7 Troubleshooting This chapter describes troubleshooting including error messages and remedies This chapter includes Error Messages Troubleshooting Initializing Maintenance...

Page 92: ...lue load exceeded the lower limit The pan or the pan base may not be properly set Check them paying attention to whether they are in contact with an external object If the error message does not disap...

Page 93: ...inimum unit weight and the percentage weighing weight limit At addition operation you placed additional samples on twice At addition operation you unloaded some samples or you pressed the key without...

Page 94: ...he batteries are exhausted Replace the batteries Display flickers The balance may be affected by an external influence such as wind and vibration Increase the setting values of relevant functions by r...

Page 95: ...ign flashing When the Zero Tare key is pressed during a sampling in parts counting mode etc The balance may be affected by an external influence such as wind and vibration The balance may be affected...

Page 96: ...dition of 1 200 bps 8 bit data 2 bit stop bit and no parity by referring to the operation manual of the external device Now output is done once after stabilization after the P key is pressed Check out...

Page 97: ...the function initialization Select by pressing the Function key Select 1 by pressing the Zero Tare key 3 Save the setting Press the Set key The balance goes back to normal display All function setting...

Page 98: ...e main unit use a piece of dry soft cloth If the unit is very dirty use a cloth soaked in a small quantity of neutral detergent or cleaning solvent If the unit is extremely dirty wash it with water an...

Page 99: ...xes Appendixes The appendixes provide data including each type of data for the balance This chapter includes Function Setting List Measurement Mode List Printing in Compliance with ISO GLP GMP Specifi...

Page 100: ...icate results Addition operation A star denotes an initial value Water Not water Only a specific gravity value Specific gravity weight actual water temperature or medium density Output disabled Auto o...

Page 101: ...put Output continuously at all times Output continuously if stable stop output if unstable Output once when the Print key is pressed whether the balance is stable or unstable Output once when the bala...

Page 102: ...le for continuous use Enable The power is turned off five minutes later Enabled only when the balance is powered by batteries Disable for continuous use Enable The backlight goes off if the balance is...

Page 103: ...ately after the key is pressed Reset the display to 0 when the balance is stabilized Disabled Enabled Off On Beeps when a value is confirmed wrong operation is made etc Beeps when the key is pressed i...

Page 104: ...only when addition function is selected 1 Parts counting Pcs 2 Cumulative count Pcs Cumulative value Displayed only when addition function is selected 3 Average unit weight Unit A Pcs Parts counting...

Page 105: ...rinting in Compliance with ISO GLP GMP 91 Appendix 3 Printing in Compliance with ISO GLP GMP Span adjustment with external weight Span test with external weight English Japanese katakana English Japan...

Page 106: ...t data Header Measurement data Footer The date and time shown in printing examples are printed only when the dedicated printer CSP 160 or CSP 240 is used Caution English Japanese katakana English Japa...

Page 107: ...late function Net addition Limit 3 point scale judgment with upper and lower limit setting absolute deviation value judgment Unit converting ISO GLP GMP compliant Tare storing Minimum readability swit...

Page 108: ...main unit CJ 220E to CJ 820E Approximately 2 kg CJ 2200E to CJ 15KE Approximately 2 5 kg Operating temperature humidity Temperature 0 to 40 C Humidity 80 rh or less No condensation allowed Options CJ...

Page 109: ...0 00005 0 00005 0 0005 0 0005 0 0005 0 0005 7 10 19 26 70 100 190 260 ozt 0 0005 0 0005 0 0005 0 0005 0 005 0 005 0 005 0 005 140 200 390 520 1400 2000 3900 5200 dwt 0 01 0 01 0 01 0 01 0 1 0 1 0 1 0...

Page 110: ...15 kg 0 001 75000 ct 5 520 oz 0 05 33 lb 0 005 480 ozt 0 05 9600 dwt 1 Bottom right grain 400 Hong Kong 0 05 390 Singapore Malaysia 0 05 400 Taiwan 0 05 4000 mom momme 0 5 1200 to 0 1 The view of the...

Page 111: ...77 0 08267 1 20565 24 11306 1 mom 3 75 18 75 0 13228 0 00827 0 12057 2 41131 1 to 11 66380 58 31902 0 41143 0 02571 0 37500 7 5 unit Grain tael Hong Kong tael Singapore Malaysia tael Taiwan momme Tola...

Page 112: ...zing 83 Interface 66 ISO GLP GMP compliant form 62 L Limit value 39 Liquid density 31 M Minimum readability 53 Momme 52 N Net 15 Net addition 36 Number of samples 21 O Operation key 5 P Pan 2 8 84 Pan...

Reviews: