OLS-85/OLS-85P
65
OLS-85
SPECIFICATIONS
12 S
PECIFICATIONS
Output power range
(typical)
1
Single port instruments
Dual port instruments
Signal stability
2
Modes
3
1 CW signal, T = 23 °C ±3 K, including connector, depending on quality of
connector applied to the OLS
2 wavelengths instruments
Laser (CW)
Laser (Modulated, Auto
)
-3.00 to 0.00 dBm (± 1 dB)
-6.00 to -3.00 dBm (± 1 dB)
3 & 4 wavelengths instruments
Laser (CW)
Laser (Modulated, Auto
)
-6.00 to -3.00 dBm (± 1 dB)
-9.00 to -6.00 dBm (± 1 dB)
SM port
Laser (CW)
Laser (Modulated, Auto
)
-3.00 to 0.00 dBm (± 1 dB)
-6.00 to -3.00 dBm (± 1 dB)
MM port
LED (CW)
LED (Modulated, Auto
)
-25.00 to -22.00 dBm
-28.00 to -25.00 dBm
2 At ambient temperature range -10 °C to +55 °C,
T = ±0.3 K
Short term
± 0.02 dB (within 15 min.)
Long term
± 0.20 dB (within 8 h)
3 Signal coding for automatic power meter wavelength detection. Works only
with Viavi power meters e.g. OLP-85
Continuous wave (CW), Modulated (270 Hz,1 kHz, 2 kHz), Auto