OLS-85/OLS-85P
19
U
NPACKING
THE
INSTRUMENT
3 G
ETTING
S
TARTED
NOTE:
Take care to insert the batteries correctly. The correct polarity is
indicated by a diagram inside the battery compartment.
– or –
Insert new or replace the RBP2 Li-Ion Battery Pack.
– or –
Switch from non-rechargeable batteries to rechargeable ones
by replacing the battery tray with a new battery pack (or vice
versa).
3.
Close the battery compartment.
4.
Press the
[
E
]
key to switch on.
Recharging the batteries
The rechargeable RBP2 Li-Ion Battery Pack is recharged when
the PS4 Universal AC/DC Power Supply is being used to power
the instrument. The instrument switches to trickle charging
automatically as soon as the RBP2 Li-Ion Battery Pack is fully
charged.
NOTE:
Rechargeable AA batteries will not be recharged in the
instrument. For AA-type rechargeable batteries please use an
external charger.
It is not possible to charge the rechargeable AA batteries or the
RBP2 Li-Ion Battery Pack via the USB interface.
The instrument is powered by the PS4 Universal AC/DC Power
Supply if the PS4 Universal AC/DC Power Supply and the USB
interface are both connected.
Deep discharge
A rechargeable battery that appears to be dead (unit will not
turn on even when connected to external power) may well be in
a deep discharge state. It can be fully charged with the following
charging cycle:
1.
Plug in the power cord for 1–1.5 hours.
Battery should be partly charged now.
2.
Switch the power on.
If the instrument starts: Let the battery fully charge.
– or –
If the instrument does not start:
Unplug power cord, plug it back in and repeat steps 1 and 2.
NOTE:
If the battery is completely deep-discharged, this cycle might
need to be repeated up to 3 times.