System Setup
4-9
Table 4-1 Setup Menu Alarms and Parameters
Menu Item
Range
Adjustment
Increment
Description
Audible Breath
ON/OFF
The breath-to-breath detection feature that can
provide an audible tone for each detected breath.
Apnea Breath
Volume
0-100
10 units
Adjusts the volume of the breath to breath
sounds.
Apnea Alarm
ON/OFF
Monitors the last breath detected. Activates an
alarm if the time between breaths exceeds the
alarm parameter.
Apnea Alarm Time
5-30 sec.
5 sec.
Times the duration between the patient’s
breaths. Activates an alarm if the set time is
violated.
High Pressure
Alarm
10-60 cmH
2
O
1 cmH
2
O
Monitors patient’s lung pressure. Activates an
alarm if the set pressure is violated.
Continuing
Pressure Alarm
10-60 cmH
2
O
1 cmH
2
O
Monitors patient’s lung pressure that is
continuous or has no indication of sufficient
exhalation. Activates an alarm if the parameter is
violated.
Screen Brightness
1-10
1 unit
Adjusts the brightness of the display.
Iso Med. Conc.
0.00% - 5.00%
0.10%
Settings to monitor the anesthetic agent output
for percentages and duration by parameters.
Each anesthetic agent has a medium and high
percentage setting and two duration settings.*
Activates an alarm if a parameter is violated.
Iso Med. Time
0-10 min.
1 min.
Iso. High Conc.
0.00% - 5.00%
0.10%
Iso High Time
0-10 min.
1 min.
Sevo. Med. Conc.
0.00% -8.00%
0.10%
Sevo. Med. Time
0-10 min.
1 min.
Sevo. High Conc.
0.00% - 8.00%
0.10%
Sevo. High Time
0-10 min.
1 min.
* Example:
To set delivery of 7% Sevoflurane for no longer than 3 minutes and deliver 5% Sevoflurane for no longer
than 10 minutes, the user can set both parameters via the
Sevo High Conc
and
Sevo Med Conc
parameters and the time duration for each.
Summary of Contents for EX3000
Page 1: ...P N 410 125 EX3000 Veterinary Anesthesia System Operator s Manual ...
Page 8: ......
Page 20: ......
Page 42: ......
Page 56: ......
Page 60: ......
Page 67: ...Replacement Parts and Accessories 10 3 Figure 10 1 EX3000 Anesthesia System ...
Page 69: ...Replacement Parts and Accessories 10 5 Figure 10 3 Optional Keyfill Vaporizer Fillers ...
Page 70: ...10 6 Replacement Parts and Accessories Figure 10 4 410 105A Isoflurane Chamber Assembly ...
Page 71: ...Replacement Parts and Accessories 10 7 Figure 10 5 410 106A Sevoflurane Chamber Assembly ...
Page 72: ...10 8 Replacement Parts and Accessories Figure 10 6 410 102A Sub Chassis Assembly ...
Page 73: ...Replacement Parts and Accessories 10 9 Figure 10 7 410 108A Circuit Board Assembly ...
Page 74: ...10 10 Replacement Parts and Accessories Figure 10 8 410 100A Back Cover Assembly ...
Page 75: ...Replacement Parts and Accessories 10 11 Figure 10 9 410 101A DISS Inlet Assembly ...
Page 76: ...10 12 Replacement Parts and Accessories Figure 10 10 500 100A Pole Assembly ...
Page 77: ...Replacement Parts and Accessories 10 13 Figure 10 11 510 570A APL Pop off Valve Assembly ...
Page 78: ...10 14 Replacement Parts and Accessories Figure 10 12 510 103A Manifold Block Assembly ...
Page 79: ...Replacement Parts and Accessories 10 15 Figure 10 13 510 101A Absorber Canister Assembly ...
Page 80: ...10 16 Replacement Parts and Accessories Figure 10 14 510 129A Flush Valve Assembly ...
Page 82: ......
Page 84: ...A 2 Appendix A Figure A 2 EX3000 Pneumatic Block Diagram EX3000 Pneumatic Block Diagram ...
Page 85: ...Appendix A A 3 Figure A 3 EX3000 Electrical Block Diagram EX3000 Electrical Block Diagram ...
Page 86: ...A 4 Appendix A Figure A 4 EX3000 Electrical Connections EX3000 Electrical Connections ...
Page 87: ......