Introduction
1-3
Facility Requirements
Table 1-1 provides the facility requirements necessary to ensure reliable operation and safety of the
EX3000 Anesthesia System.
Table 1-1 Facility Requirements
Characteristic
Specification
Working Area
Large enough to accommodate the anesthesia system. See Figure 1-2.
Power
95 - 240 VAC, 50/60 Hz
Note:
An AC power fail alarm sounds with both audible and visual indicators if
power supply drops below 95 VAC. The alarm is self-resetting.
Oxygen Supply
Delivery Options:
x
Oxygen supply hose from the facility oxygen outlet
x
Type E cylinder mounted to the anesthesia machine via an optional
manifold and yoke assembly
x
Type H cylinder strapped to facility wall or a cylinder cart
Regulated Inlet Pressure: 50–60 psi (345–380 kPa)
Note:
A low O
2
pressure alarm sounds with both audible and visual indicators
if supply pressure drops below approximately 45 psig. Alarm is self-resetting.
Flow Rate: 20 liters per minute at 55 psi (379 kPa)
Oxygen: Medical Grade
Pneumatic Fitting
Diameter Indexed Safety System (DISS) for Oxygen
Environment
Storage Temperature: 41 to 122
q
F (5 to 50
q
C)
Operating Temperature: 59 to 95
q
F (15 to 35
q
C)
Relative Humidity: 5 to 90% (non-condensing)
Altitude: 1609 meters (5278 feet) maximum
Note:
If you have any questions about facility requirements, please contact
Vetland
or your
authorized distributor.
Summary of Contents for EX3000
Page 1: ...P N 410 125 EX3000 Veterinary Anesthesia System Operator s Manual ...
Page 8: ......
Page 20: ......
Page 42: ......
Page 56: ......
Page 60: ......
Page 67: ...Replacement Parts and Accessories 10 3 Figure 10 1 EX3000 Anesthesia System ...
Page 69: ...Replacement Parts and Accessories 10 5 Figure 10 3 Optional Keyfill Vaporizer Fillers ...
Page 70: ...10 6 Replacement Parts and Accessories Figure 10 4 410 105A Isoflurane Chamber Assembly ...
Page 71: ...Replacement Parts and Accessories 10 7 Figure 10 5 410 106A Sevoflurane Chamber Assembly ...
Page 72: ...10 8 Replacement Parts and Accessories Figure 10 6 410 102A Sub Chassis Assembly ...
Page 73: ...Replacement Parts and Accessories 10 9 Figure 10 7 410 108A Circuit Board Assembly ...
Page 74: ...10 10 Replacement Parts and Accessories Figure 10 8 410 100A Back Cover Assembly ...
Page 75: ...Replacement Parts and Accessories 10 11 Figure 10 9 410 101A DISS Inlet Assembly ...
Page 76: ...10 12 Replacement Parts and Accessories Figure 10 10 500 100A Pole Assembly ...
Page 77: ...Replacement Parts and Accessories 10 13 Figure 10 11 510 570A APL Pop off Valve Assembly ...
Page 78: ...10 14 Replacement Parts and Accessories Figure 10 12 510 103A Manifold Block Assembly ...
Page 79: ...Replacement Parts and Accessories 10 15 Figure 10 13 510 101A Absorber Canister Assembly ...
Page 80: ...10 16 Replacement Parts and Accessories Figure 10 14 510 129A Flush Valve Assembly ...
Page 82: ......
Page 84: ...A 2 Appendix A Figure A 2 EX3000 Pneumatic Block Diagram EX3000 Pneumatic Block Diagram ...
Page 85: ...Appendix A A 3 Figure A 3 EX3000 Electrical Block Diagram EX3000 Electrical Block Diagram ...
Page 86: ...A 4 Appendix A Figure A 4 EX3000 Electrical Connections EX3000 Electrical Connections ...
Page 87: ......