26
■
Operation
Operation Procedure
Caution
REMEMBER
Operator and people who stand near by the operator shall be wear ear
protector (e.g. earmuff). Long time exposure under so larger sound will cause
hearing impairment.
REMEMBER
Perform daily check before putting in use. G-9 shall be put in use with
monitoring or management system in case of that performance of G-9 has
great impact to your work.
Perform daily check before putting in use. See the table below as basic check points and add
check points to fit your managing requirements. This check points do not guarantee all
performance and functionalities of G-9.
Parts
Inspection
Judge
Body, every parts of body,
motor cartridge, nozzle.
Visual examination
No cracks, chips, distortions,
measure scratch and
discoloration.
Body, every parts of body,
motor cartridge, nozzle.
Vibrate a little by hand.
No large clattery sound from
inside.
No loosening.
Seam of body, trigger, motor
cartridge, air fitting
Air leak test by hearing and / or
touching
No air leaking.
Regulator
Loosen examination
No rotate with light twisting.
Motor cartridge
Loosen examination
No rotate with light twisting.
Nozzle
Loosen examination
No rotate with light twisting.
Performance
Test piece and static voltage
tester (VESSEL
Eye-02
)
Has enough performance.
Ion balance
Static voltage tester (VESSEL
Eye-02
)
<
±
30V
Trigger
Grab and release without air
supply
Back to the place before until 1
second.
Grab the trigger then ionized compressed air blow from nozzle.
Put G-9 in use always with fully grabbing the trigger. Adjust compressed air pressure and
regulator alternately to get appropriate air flow.
There may be small time rag from grab the trigger to perform ionizing. This is not a malfunction.