Function Settings
Measurement mode selection
“Static charge voltage measurement” and “Ion Balance Measurement”
When the STATIC/I.B. button is pressed after starting up the device, the mode can be switched
between “Static charge voltage measurement mode” and “Ion Balance measurement mode”.
Static charge voltage
measurement mode
Ion Balance voltage
measurement mode
Mode Display
IB
The “Static charge voltage
measurement mode” is active when
the power is turned on.
Press the STATIC/I.B. button after
turning the power on.
* Press again to return to the “Static
charge voltage measurement mode”.
LCD display
Alarm
No
No
* Attach the Ion Balance plate before measuring the Ion Balance. Refer to “Handling the Ion
Balance plate” for information on attaching the plate.
Zero adjustment
If the ZERO button is held down after starting up the device, the measurement value and bar graph
are cleared to zero.
Zero adjustment
* This functions only when the display
value is within the tolerable range of
digital zero operation.
(Static charge voltage: within ± 0.5 kV)
(Ion Balance: within ± 100 V)
* This does not function if the
measurement value is fixed with the
HOLD/MAX value.
* Note that if the ZERO button is
inadvertently pressed during
measurement, the digital zero
adjustment will activate and the zero
point will move.