Ion Balance Measurement
Continuous mode for Ion Balance measurement
1) Press the power button for 3.5 to 4.5 seconds. When the power turns on and “3” appears on the
LCD display, release the button. A “beeping” sound is made to indicate that the “auto power off
function” has been canceled, and the “continuous measurement mode” has been entered. The
bar graph, scale, display range, polarity, measurement value, unit (kV), battery level and “A.ON”
appear on the LCD screen. The red LED for the cross mark turns on.
2) Next, press the STATIC/I.B. button once. The IB mode appears, the range and unit (V) change,
and the red LED for the cross mark disappears.
Always check the battery level at this time. Refer to “Battery level display” for details on the
battery level.
3) Lightly touch the front of the Ion Balance plate with a finger to release the charge. (If wearing
gloves, touch with a grounded wire, etc.) If a value other than 0 V appears on the LCD screen,
press the ZERO button to adjust the zero point. (Refer to page 9.)
4) When the power button is pressed again in the power on state, a single “beep” sounds, the
power turns off, and all displays turn off.
Continuous mode for Ion Balance measurement
REMEMBER
Eye-02 cannot be turned off without pressing POWER button in
continuous measurement mode. Confirm that power is turned off by
observing the disappearance of all indicators. If power is not turned off,
the battery will be drained completely.
* Eye-02 cannot return to this mode automatically, even if it is turned on again after turning off in
continuous mode. For going into continuous Ion Balance measurement mode, the operation
described in “Continuous mode for Ion Balance Measurement” should be carried out when power
is turned on.