Ion  Balance  Measurement 

Continuous mode for Ion Balance measurement

 

1) Press the power button for 3.5 to 4.5 seconds. When the power turns on and “3” appears on the 

LCD display, release the button. A “beeping” sound is made to indicate that the “auto power off 

function” has been canceled, and the “continuous measurement mode” has been entered. The 
bar graph, scale, display range, polarity, measurement value, unit (kV), battery level and “A.ON” 

appear on the LCD screen. The red LED for the cross mark turns on. 

2) Next, press the STATIC/I.B. button once. The IB mode appears, the range and unit (V) change, 

and the red LED for the cross mark disappears. 

Always check the battery level at this time. Refer to “Battery level display” for details on the 

battery level. 

3) Lightly touch the front of the Ion Balance plate with a finger to release the charge. (If wearing 

gloves, touch with a grounded wire, etc.) If a value other than 0 V appears on the LCD screen, 

press the ZERO button to adjust the zero point. (Refer to page 9.) 

4) When the power button is pressed again in the power on state, a single “beep” sounds, the 

power turns off, and all displays turn off. 

 

 

 

Continuous mode for Ion Balance measurement 

 

REMEMBER

 

Eye-02 cannot be turned off without pressing POWER button in 

continuous measurement mode. Confirm that power is turned off by 

observing the disappearance of all indicators. If power is not turned off, 

the battery will be drained completely. 

* Eye-02 cannot return to this mode automatically, even if it is turned on again after turning off in 

continuous mode. For going into continuous Ion Balance measurement mode, the operation 
described in “Continuous mode for Ion Balance Measurement” should be carried out when power 

is turned on. 

 
 

 

 
 

 

 

Summary of Contents for EYE-02

Page 1: ...e Keep it in a safe handy place for future reference Table of Contents For Your Safety P 2 Product Features P 3 Product Specifications P 4 Preparation and Battery Replacement P 6 Function Settings P 6 Static charge voltage Measurement P 7 Ion Balance Measurement P 12 Table of Functions P 18 Troubleshooting P 19 Warranty P 20 ...

Page 2: ...O NOT WET Do not use the Eye 02 in highly damp places or anywhere it may be wet by water oil solvent or other liquid Contact with moisture can result in electric shock or breakdown PROHIBITED Do not press on the LCD display from the top PROHIBITED Do not apply any mechanical vibrations and shock PROHIBITED Do not insert any object in the opening for the sensor located at the measuring side of the ...

Page 3: ...e The maximum value peak value can be displayed by switching to the MAX mode Beep sound Beep sound will be heard during Power on Power off Over range etc Cross mark indicated by red LED The Static charge voltage can be measured with 22 KV 22 000 V at a distance of 25 mm The cross mark indicated with the red LED shows the proper distance for a clear view Ion Balance Measurement The Ion Balance volt...

Page 4: ...s buzzer sounds when 220 V is exceeded Digital reading 0 to 220 kV 000 V Bar graph display 15 V per bar Accuracy 10 LCD display renewal rate 5 times second Response time less than one second Auto power off Power is turned off automatically after five minutes Switch between on and off by pressing the power button Display lamp LED white lamp to assist reading in insufficient light Switch between on ...

Page 5: ...5 Product Specifications Accessories Ion Balance PLATE 1 Ground Cord 1 With alligator clips Strap 1 Soft Case 1 Overview and Name of parts ...

Page 6: ...all the battery with the following procedure 1 Press down on the indent on the back lid and slide the lid open 2 Connect the battery to the connection plug Pay attention to the battery s polarity 3 Set the battery into the battery compartment and close the back lid Installing the battery REMEMBER Do not exert excessive pressure it might damage the lid REMEMBER The connection plug is delicate so ta...

Page 7: ...n The grounding terminal functions as the reference for the measurement circuit Always ground the terminal to ensure accurate measurements The measurement accuracy will not be guaranteed if the terminal is not correctly grounded Handling the Ion Balance plate This device is equipped with an Ion Balance plate for measuring the Ion Balance The Ion Balance plate is mounted on the device s sensing sec...

Page 8: ...isplay Alarm Beep Beep Beep Beep Beep Beep Beep Beep The auto power off and backlight settings are valid in the Static charge voltage measurement mode and the Ion Balance measurement mode If the power button is held down after 4 is displayed Mode 1 will reappear The mode selection will repeat as long as the power button is held down If the power button is pressed again in the power on state the po...

Page 9: ...ement mode LCD display Alarm No No Attach the Ion Balance plate before measuring the Ion Balance Refer to Handling the Ion Balance plate for information on attaching the plate Zero adjustment If the ZERO button is held down after starting up the device the measurement value and bar graph are cleared to zero Zero adjustment This functions only when the display value is within the tolerable range of...

Page 10: ...e During normal measurement the display will fluctuate when the product is moved When hold the maximum value during measurement MAX is selected the maximum value in the measurement always appears without stopping the measurement Temporarily hold measurement Hold maximum measurement value Mode Display HOLD MAX Press HOLD MAX button during measurement Press again to cancel Hold down the HOLD MAX but...

Page 11: ...button is pressed again in the power on state a single beep sounds the power turns off and all displays turn off 4 This product has an auto power off function When enabled the power turns off automatically about five minutes after the power is turned on Four beeps sound to indicate that the power has turned off Standard Static charge voltage measurement mode REMEMBER Eye 02 can be activated by pre...

Page 12: ...t state is activated Always check the battery level at this time Refer to Battery level display for details on the battery level 3 When finished with measurement turn the power button A single beep sounds the power turns off and all displays turn off Continuous mode for static charge voltage measurement REMEMBER Eye 02 cannot be turned off without pressing POWER button in continuous measurement mo...

Page 13: ...e NG The LED cross mark function is adjusted so that the cross pattern irradiated by the red LED light is clearest on a flat plate placed at a measurement distance of 25 mm This can be confirmed by facing the Eye 02 toward a piece of white paper Measurement and cross mark REMEMBER If the digital numerical value blinks at 22 0 and a continuous beeping sound is heard while moving the measuring devic...

Page 14: ...polarity of the static charge red being positive and blue negative Right Positive charge Left Negative charge Each bar of the bar graph corresponds to 0 1 kV 100 V approximately for low range and 1 5 kV 1500 V approximately for high range Eye 02 is calibrated for the voltage range 0 to 20 kV using a 150 mm x 150 mm charged flat metal plate placed at a distance of 25 mm For any other distance accur...

Page 15: ...case of Eye 02 is made of conductive resin The grounding terminal provides the reference potential for the electrical circuit This terminal should be grounded for proper measurement The accuracy is not guaranteed for measurement with an ungrounded Eye 02 Tripod attachment A tripod can be attached to the camera threads female threads on the bottom of the case Tripod must be prepared separately The ...

Page 16: ... with a grounded wire etc If a value other than 0 V appears on the LCD screen press the ZERO button to adjust the zero point Refer to page 9 4 When the power button is pressed again in the power on state a single beep sounds the power turns off and all displays turn off 5 This product has an auto power off function When enabled the power turns off automatically about five minutes after the power i...

Page 17: ...ouch the front of the Ion Balance plate with a finger to release the charge If wearing gloves touch with a grounded wire etc If a value other than 0 V appears on the LCD screen press the ZERO button to adjust the zero point Refer to page 9 4 When the power button is pressed again in the power on state a single beep sounds the power turns off and all displays turn off Continuous mode for Ion Balanc...

Page 18: ... into Ion Balance measurement mode when Ion Balance button is pressed even if the Ion Balance plate is not attached or not attached properly However the measured value will not be correct Eye 02 is calibrated by applying DC voltage to the Ion Balance plate directly If the plate is attached properly to EyeE 01 the display gives a direct reading of the Ion Balance voltage Ion Balance Measurement REM...

Page 19: ...nded metal Eye 02 is turned toward a charged object with bad grounding or the operator is charged Turn toward an uncharged object Ground Eye 02 using the grounding cord Display shows Err Malfunction of the sensor Repair or replacement of the sensor is needed Beep sound does not activate Malfunction of the sensor Repair or replacement of the buzzer or PCB is needed Indicated voltage is high in Ion ...

Page 20: ...idents neglect misuse alterations operator error or failure to properly maintain clean or repair products LIMIT OF LIABILITY This electronic static eraser use high voltage corona discharge and should not be used in or near flammable or explosive environments In no event will VESSEL or any seller is responsible or liable for any injury loss or damage direct or consequential whether based in tort or...

Reviews: