13
■
Static charge voltage Measurement
Measurement and cross mark
Move the sensing section on the front of the
device gradually toward the charged object to be
measured. This Eye-02 has an LED cross mark
function which provides a guide of the distance
from the charged object. Move the Eye-02 toward
the charged object until the irradiated cross mark
is clear.
Measurement distance
←
Far (NG)
←OK
←
Close (NG)
* The LED cross mark function is adjusted so that the cross
pattern irradiated by the red LED light is clearest on a flat
plate placed at a measurement distance of 25 mm. This
can be confirmed by facing the Eye-02 toward a piece of
white paper.
Measurement and cross mark
REMEMBER
If the digital numerical value blinks at [
±
22.0] and a continuous
“
beeping
”
sound is heard while moving the measuring device, the
charged voltage exceeds the measurement range (over range) of this
measuring device, so stop the measurement. The internal sensor could be
damaged if measurements are continued in this state.