17.3
Calibrate
To calibrate the balance a 5 kg calibration mass is required. The balance must be
calibrated after installation and if the equipment is moved it is also good working
practice to check calibration periodically. This should be done with the chamber at room
temperature and with the afterburner fan switched off. To start the calibration process,
place the calibration plate (See Fig 8.) onto the balance extension press the calibrate
button. The status message sequence for normal operation is:
l
Clear pan then press Next
l
Load 5 kg then press Next
l
Calibration Complete
After pressing the calibration button the ‘Next’ and ‘Cancel’ buttons appear. ‘Cancel’ can
be pressed at any time to abort the operation. These buttons can be seen in Fig 68.
NOTE: When loading 5 kg the weight reading will remain at zero, this is normal. After
the calibration is complete the balance weight will read 5000 g, until the weight is
removed.
NOTE: After calibration, remove the calibration plate. This will result in a negative
weight value showing in the display. Restart the furnace using the main power switch
in order to clear this value and commence normal operation.
Fig 68 - Screen-view: Balance Control Screen - Calibration
65
17.0 Internal Balance
Summary of Contents for CARBOLITE GERO ABA 7/35B
Page 11: ...2 2 Identifying the Parts 11 2 0 Supplied Item List ...
Page 19: ...Fig 14 Schematic for Mains Electrical Connections 19 3 0 Installation ...
Page 57: ...Fig 61 Screen view Standard Recipe Fig 62 Calibration Recipe 57 14 0 Recipe Editor ...
Page 70: ...20 0 Temperature Calibration Fig 72 Screen view Dual Point Calibration 70 ...
Page 93: ...Notes Service Record Engineer Name Date Record of Work ...