7.0 Running a Test
Fig 39 - Screen-view: Pre-test Display
Fig 40 - Screen-view: In-test Display
During a test the controller will continuously update the values circled in green, shown in
Fig 38.
The values circled in red are updated every 30 seconds after the initial delay time +
constant weight time has elapsed. For example, if the initial delay time is set to 10
minutes in the recipe and constant weight time is set to 3 minutes, then the values will
be updated after 13 minutes and then every 30 seconds until the end of the test. These
values indicate the weight loss during previous ‘x’ minutes, where ‘x’ is the constant
weight time. If the recipe type is ‘calibration’, the calibration factor, circled in yellow, will
be updated on test completion and if the recipe type is ‘standard’ then the binder %,
circled in blue, will be updated on test completion. If a calibration recipe has been
selected then its name will be preceded by a tilde ‘~’ in the recipe name field. The recipe
names use the same indicator in the recipe library.
The 'Test Status' field has the possible states shown in the table below.
34