5.4
Operator Safety
The ABA incorporates a safety switch which interrupts the heating
element circuit when the door is opened. This prevents the operator
from touching a live heating element, but also prevents heating up if the
door is left open.
When a test is started the door is automatically locked until a constant weight loss (end-
point) is achieved. This is to prevent the operator from opening the door whilst volatiles
are being driven off from the test sample. If the door were to be opened during this
time the addition of air could cause the volatile materials to ignite in an explosive
manner.
The red sample basket cover must be used to cover the hot sample
baskets whilst they are allowed to cool.
Before removing any hot objects from the chamber make sure that you
have a safe place to put them down.
Do not place hot objects on inflammable surfaces. Wear appropriate
safety clothing; gloves, mask and safety glasses are recommended.
5.5
Touch-screen Main Menu
The main menu is accessed directly from the button in the top left corner of the home
screen as shown in Fig 36.
The table below gives a reference to a section of this document where detailed
information is given for each of the buttons on the Main Menu.
Fig 36 - Screen-view: Main Menu
29
5.0 ABA Operation
Summary of Contents for CARBOLITE GERO ABA 7/35B
Page 11: ...2 2 Identifying the Parts 11 2 0 Supplied Item List ...
Page 19: ...Fig 14 Schematic for Mains Electrical Connections 19 3 0 Installation ...
Page 57: ...Fig 61 Screen view Standard Recipe Fig 62 Calibration Recipe 57 14 0 Recipe Editor ...
Page 70: ...20 0 Temperature Calibration Fig 72 Screen view Dual Point Calibration 70 ...
Page 93: ...Notes Service Record Engineer Name Date Record of Work ...