7.0 Running a Test
To connect an Ohaus Explorer™ balance a non-standard RS232 cable with the
connections shown in Fig 42. must be used. The cable connects the balance to the 9 way
female D-type connector at the rear of the furnace.
Fig 44 - Ohaus Explorer Serial Cable Connections
The balance settings are shown in the table below.
7.10.1
Table: Ohaus Balance Settings
Auto Print
Off
Print Internal
Off
Numeric Data
Off
Units: Gram (g)
On
Baud Rate
9600
Parity
None
Data Bits
8
Stop Bits
1
7.11
Connecting an Ohaus Explorer Pro™ External Balance
This section refers to the Ohaus balance with a colour touch screen interface.
The external balance type must be set to Ohaus Explorer™ in the 'General Settings'
screen (see Fig 45.).
A standard straight through RS232 cable is used to connect the balance to the 9 way
female RS232 connector at the rear of the furnace.
The balance settings are shown in Table: Ohaus Balance Settings in the previous section
(same settings as for Ohaus Explorer).