Page 13
Step 17 - Auto-restart (screen 033)
This allows selection of the AUTO restart option. If auto
restart is selected the unit will automatically restart after a
power failure. In manual mode the unit will stay in standby
after a power failure.
1.
PRESS
to change mode (
= ON)
OR
PRESS
✔
to proceed to step 18.
2.
PRESS
✔
.
Step 18 - Feed Water Quality (screen 034)
An indication of the RO performance can be obtained
using a calculation of ionic rejection in which the
conductivity of the permeate is compared to that of the
feedwater.
Enter the feedwater conductivity to obtain an accurate
measurement.
1.
PRESS
to reduce the value to the correct reading
2.
PRESS
✔
to accept
Step 19 - Data output (screen 035)
The
MEDICA
can transmit operational data to an Xport
storage device.
1.
PRESS
to change (
= ON)
OR
PRESS
✔
to proceed to step 20.
2.
PRESS
✔
.
Step 20 - Data transmit (screen 036)
Setting the frequency of data transmittals to Xport.
1.
PRESS
to change transmit intervals (1, 5, 15,
30min / 1,6 hour)
OR
PRESS
✔
to proceed to step 21.
2.
PRESS
✔
.
Note: PRESS PRINT during normal operation and
current data will be transmitted if data output (035)
is selected.
Auto-restart screen
Data transmit screen
Feed Water Quality Screen
Data output screen