6.
SANITIZATION
6.1
RO sanitization (Step 1)
The RO is sanitized to reduce the growth of microbiological
contamination within the RO module. The
MEDICA
has a built-in
reservoir which allows the sterilant to be introduced into the water
feeding the RO. Please read this entire section to become familiar
with the procedure before you start.
The sanitization frequency required is dependent on the feedwater,
local environment, usage patterns and application. As a general
rule, ELGA LabWater recommends sanitization of the RO at 6
monthly intervals. However, the period between sanitizations could
be extended in particular circumstances. For example, microbial
growth will usually be lower with clean feed water, well-maintained
pre-treatment, low temperatures and heavy usage.
WARNING! ALWAYS WEAR RUBBER GLOVES, APRON
AND FACEMASK. DO NOT BREATHE FUMES
OR ALLOW TO COME IN CONTACT WITH
SKIN
OR
EYES.
ALWAYS
FOLLOW
RECOMMENDATIONS
FOUND
IN
THE
MANUFACTURERS MATERIAL SAFETY DATA
SHEET
AND
ANY
APPLICABLE
OSHA
STANDARDS FOR THE CHEMICAL BEING
USED.
WARNING! LABEL THE MACHINE WITH APPROPRIATE
WARNING SIGNS SUCH A
S “DO NOT USE/
CONTAINS STERILANT” (NOT PROVIDED)
Minncare Cold Sterilant is a Peracetic and Hydrogen Peroxide
based solution.
Refer to the Minncare label for additional information and follow all
applicable directions for use on the manufacturers label in
conjunction with the following instructions.
Sanitization selection screen
1.
APPLY a suitable warning label such as “DO NOT
USE / CONTAINS
STERILANT”.
2.
ISOLATE the final application.
3.
PRESENT GREEN sanitization key to reader.
4.
PRESS
✓ to enter sanitization.