MEDICA Pro 30/60/120 US Operator Manual
ELGA
Page 10
MEDICA Pro 30/60/120 US MANU40594 Version 1 - 07/14
Step 9 - RO water quality alarm settings QS1 (screen 025)
This screen is used to set the value at which the RO water
quality alarm will sound. This alarm does not stop RO
production but can extend or start flush routines.
1.
PRESS
to select alarm point (Increments of 10,
ranging
from
20
to
100µS/cm).
If ‘0’ is selected the alarm will be turned ‘OFF’.
OR
PRESS
✔
to proceed to step 10.
2.
PRESS
✔
.
Step 10 - Product water purity alarm settings QS2 (screen 026)
This screen is used to select the value at which the product
water purity alarm will activate. The alarm does not stop
the unit and will automatically reset if the purity level
recovers
1.
PRESS
to select alarm point (increments of 1,
ranging from 1 to 10 M
.cm).
OR
PRESS
✔
to proceed to step 11.
2.
PRESS
✔
.
Step 11 - RO water temperature alarm settings TS1 (screen
027)
This screen is used to select the value at which the RO
water temperature alarm will activate. The alarm does not
stop the unit and will automatically reset if the temperature
returns below the set point.
1.
PRESS
to select alarm point (increments of 1,
ranging from 20 to 50
C)
OR
PRESS
✔
to proceed to step 12.
2.
PRESS
✔
.
Alarm settings TS1 screen
Alarm settings QS1 screen
Alarm settings QS2 screen