6 Set up with the display and adjustment module
VEGADIS 82 • 4 … 20 mA/HART
45300-EN-170724
Proceed as follows:
1. Select with
[->]
the menu item "Upper Range" and confirm with
[OK]
.
2. Edit the percentage value with
[OK]
and set the cursor to the
requested position with
[->]
.
3. Set the requested percentage value (e.g. 90 %) with
[+]
and save
with
[OK]
. The cursor jumps now to the pressure value.
4. Enter the pressure value for the full vessel (e.g. 900 mbar) cor-
responding to the percentage value.
5. Store settings with
[OK]
and move to the Lower Range with
"
ESC
".
The adjustment of the Upper Range is hence finished.
For an adjustment with filling, simply enter the actual measured value
indicated at the bottom of the display.
Proceed as follows:
1. Select with
[->]
the menu item "
Lower Range
" and confirm with
[OK]
.
2. Edit the percentage value with
[OK]
and set the cursor to the
requested position with
[->]
.
3. Set the requested percentage value (e.g. 10 %) with
[+]
and save
with
[OK]
. The cursor jumps now to the pressure value.
4. Enter the pressure value corresponding to the min. level (e.g.
0 mbar).
5. Save settings with
[OK]
and move with
[ESC]
and
[->]
to the max.
adjustment.
The adjustment of the Lower Range is hence finished.
To damp process-dependent measured value fluctuations, set an inte-
gration time of 0 … 999 s in this menu item. The increment is 1 s.
In this menu item, the device status is displayed.
Setup - Upper Range
Setup - Lower Range
Setup - Damping
Diagnostics - Device
status