66
9 Diagnostics and servicing
SOLITRAC 31 • Profibus PA
43834-EN-211203
Code
Text message
Cause
Rectification
F066
Faulty adjustment
Adjustment not yet carried out
Error during adjustment or when enter-
ing the linearisation table
Carry out adjustment
Carry out linearisation
F068
Count rate too high
Faulty instrument settings (1031)
Faulty steam density compensation
(1101)
Carry out a reset
Check Secondary instrument (steam
density)
F072
Limit exceeded
Faulty instrument settings
Carry out a reset
F073
Error real value correction
Faulty real value correction
Carry out real value correction again
F080
System error
Instrument error
Restart instrument
Call our service
F086
Communication error
Error in the Fieldbus communication
Restart instrument
Call our service
F114
Error real time clock
Discharge accumulator
Readjust real time clock
F120
Filter time error
Faulty or missing instrument adjustment Carry out adjustment
F121
Faulty participant list on the
multisensor communica-
tion bus
Secondary instruments not found
Secondary instrument with wrong ad-
dress
Check Secondary instruments
Check Secondary list in Primary instru-
ment
Correct address of Secondary instru-
ments
F122
Double addresses on the
multisensor communica-
tion bus
Instrument addresse was assigned sev-
eral times
Change instrument addresses
F123
X-ray alarm
External instruments cause radiation
Radiation above the max. adjustment
value
Determine reason for X-ray alarm
In case of brief X-ray radiation: Monitor
switching outputs for this time manually
F124
Alarm due to increased ra-
diation
Radiation dose too high
Determine reason for increased radia-
tion
F125
Ambient temperature too
high
Ambient temperature on the housing
outside the specification
Cool the instrument or protect it with iso-
lating material against radiation heat
F126
Error in the trend recording
Instrument error
Call our service
F141
Communication error on the
multisensor communica-
tion bus
Secondary instrument does not answer Check Secondary instruments
Tab. 4: Error codes and text messages, information on causes as well as corrective measures