Distance range (km, m, miles, kilofeet, ft):
maximum value of measured distances;
the value of distance range should exceed the expected length of the fiber under test
(20% more). The distance unit type can be set by clicking the
Display
button (FX150) or
selecting the unit in the
Distance Unit
drop-down (RXT-4111/RXT-4113).
•
Pulse Width, ns
: the value of the optical pulse duration; the range of available optical
pulse values depend on the selected distance range and is set by the OTDR software
automatically (10 ns pulse ~ 1 meter). Short pulse widths provide better resolution and
produce the best event and attenuation dead zones. Short pulse widths, however, inject
less optical energy into the Fiber-Under-Test (FUT), resulting in a noisier trace and
lower dynamic range. Longer pulse widths provide the highest dynamic range and are
used to test long fibers.
•
Resolution, m
: distance between two data points (sampling interval) of the trace;
values depend on the selected distance range. Select smallest value to measure events
spaced close together and use largest value to for longer ranges. Auto mode will deter-
mine best value for Fiber Under Test.
•
Time, min:sec
: determines the duration of measurement process with averaging. Use
Live mode to perform monitoring of the fiber in Real Time. Longer averaging time
improves the SNR and produces a “smoother” trace.
•
Optimized For:
test modes - Automatic DR/DZ, maximize dynamic range (DR mode) or
best deadzone, sacrifice range (DZ mode)
7.2.2 Auto Mode
To set OTDR test parameters automatically, under
Test Parameters
select
Auto
from the
Mode
drop-down box. In this mode, the OTDR performs a measurement and analysis of the
fiber and determines optimal test settings automatically.
OTDR Setup: FX150 Auto mode
Summary of Contents for FX150
Page 14: ...OTDR Series e Manual RevE00 Page 14 of 128 TX3x0S Top View RXT 4100 OTDR Module ...
Page 17: ...OTDR Series e Manual RevE00 Page 17 of 128 5 3 MTTplus Overview 5 3 1 MTTplus Side View ...
Page 19: ...OTDR Series e Manual RevE00 Page 19 of 128 5 4 RXT Overview ...
Page 50: ...OTDR Series e Manual RevE00 Page 50 of 128 Analysis Report pdf format ...
Page 117: ...OTDR Series e Manual RevE00 Page 117 of 128 Results tab Wavelength Trace ...