2. Choosing an Installation Site
PaX-i Plus / Insight Installation Manual
11
2.
Choosing an Installation Site
2.1
Room Requirements
The location of this equipment should allow for high visibility of
the patient by the operator and the operator should be as near
to the patient as possible.
This equipment should not be installed on thick carpets for
stability reasons.
Anti-static floor materials should be used around the
equipment.
The PC monitor, emergency cut off switch and X-ray
Exposure Switch
should be installed in the vicinity of the
operator so that he or she can manage them simultaneously in
case of emergency.