4. Installing the Equipment: Base Stand (Optional)
60
PaX-i Plus / Insight Installation Manual
4.9
Leveling the Equipment
T-shaped Hex
Wench
8 mm / 0.3”
Sprit Level
n/a
Ensure that the Spirit Level should restonly on the locations
indicated in the following figures to obtain the accurate center.
1.
Prepare the Spirit Level.
2.
Turn the Rotating Unit clockwise so that the X-ray tube head faces the front as shown
in the figure.
3.
Remove the case column front.
4.
Turn all screws on the base plate unit clockwise until they touch the ground.