4. Installation Procedures
50
Green21 (Model: PCT-90LH) Installation Manual
2) Before inputting position correction value, [SPM_CLI?] & [SPM_CRI?] Check
the value set by the command first and add [SPM_CLIC_xxxx] &
[SPM_CRIC_xxxx] by adding the calculated value to the set value.
ex) Check setting value: [SPM_CLI?]
0006(=+6)
Calculation of input values:
+39 (calculated value) + +6 (set value) = +45
Input [SPM_CLIC_0045]
If the calculated value is negative, enter 1 for the first digit.
3) Run C:\VCaptureSW\Exe\VCaptureSW.exe file and select FOV 21x19 then
shoot (No Reconfiguration Required).
When performing the collimator correction, wait for 10 minutes
after performing one CT sequence shot and then take another
shot.
4) Run C:\VCaptureSW\Util\EzEval_Phantom\EzEval.exe file and final check if
the criteria value is satisfied. (If out of the criteria, rework in the order of 1 to
2.)
5) If the criteria value is satisfied, backup the corresponding PROJ folder. (Ex:
PROJ_Init)
Collimator speed and start position compensation are only
performed by CT, and the Pano is automatically synchronized
when the CT operation is completed.