4. Installation Procedures
46
Green21 (Model: PCT-90LH) Installation Manual
4.10 Check X-ray align (4% Cut)
Make sure to remove the inner Carbon Cover of the Rotating
Unit before conducting this section.
* Disassembly of the inner Carbon Cover.
Make sure the X-ray align is correct. The criteria for left, right, top and bottom are
as follows.
1) CT: Threshold - 95
Collimator
Min
Max
Left (T-80%)
6 pixels
8 pixels
Right (T-80%)
4 pixels
6 pixels
Top (T-80%)
4 pixels
10 pixels
Bottom (T-80%)
4 pixels
10 pixels
2) Pano: Threshold - 80
Collimator
Min
Max
Left (T-80%)
3 pixels
5 pixels
Right (T-80%)
2 pixels
4 pixels
Top (T-80%)
275 pixels 290 pixels
Bottom (T-80%)
15 pixels
30 pixels