2. Choosing an Installation Site
8
Green21 (Model: PCT-90LH) Installation Manual
Ceiling Height: ≥ 2,500mm/98”
Dimension (mm/inch)
2,500mm(L) x 2,000mm(W) x 2,500mm(H) / 98”(L) x 79”(W) x 98”(H)
If the ceiling height is less than 2,285 mm (= max. height of the
Column Unit + 100 mm), refer to
‘Appendix B. Limiting the
on
to lower the max. height of
the Column Unit.
2. The width of the entrance:
The door of the X-ray room should have a clearance of more than 800 mm (31”)
wide.
3. Floor area:
The floor of the X-ray room must be stable and level for system balance. The floor
must be able to support a minimum weight of 500 kg/m
2
(110 lbs/feet
2
).
4. Protection against radiation:
To protect against radiation hazards, follow all federal and municipal
requirements.
During exposure, the operator should follow applicable radiation shielding
requirements and remain at least 2m (7’) from the source of the radiation.
Maintain visible contact with the patient and a clear view of indicators such as the
warning lamp and imaging status on the PC.