Shut Down
As a general rule, it is recommended that the flowcell be flushed
with several volumes of clean, non-ionic eluant. This is
especially important if ionic buffer solutions have been used.
After flushing, simply turn the power switch on the back panel to
the OFF (downwards) position.
The lamp may be shut-off remotely with the rear panel located
Lamp Shut-off terminal. This will prolong lamp life while the
detector is not in use. The lamp is shut off by providing a
continuous contact closure or TTL low. The lamp is re-ignited by
interrupting the contact closure or TTL low.
NOTE:
After the lamp is remotely re-ignited, the strip chart recorder output will
transmit a voltage related to the current flowcell absorbance and the last
stored zero value in the autozero circuit