S
YSTEM
D
ESCRIPTION
ProStar 340 UV/Vis Detector
7
•
Pressing the autozero switch. After the new zero value
has been stored, the light turns off and the recorder
outputs will become active.
9. Event Switch: A momentary switch sends an event mark of
approximately 20% deflection to the recorder output. This
switch does not affect the integrator output.
10. Short Switch: A momentary switch shorts the recorder
output terminals to zero volts. Pressing and holding the
switch allows the user to set the chart recorder pen position.
This switch does not affect the integrator output.
11. Reference Light Intensity Switch: A momentary switch
displays a value which is proportional to the light intensity at
the reference photodiode.
12. Sample Light Intensity Switch: A momentary switch displays
a value which is proportional to the light intensity at the
sample cell photodiode.
13. Visible (VIS) Lamp Indicator: A green LED indicates that a
Tungsten lamp is present in the instrument and is lit.
14. Ultra Violet (UV) Lamp Indicator: A green LED indicates that
a Deuterium (D2) lamp is present in the instrument and is lit.
1
2
3
4
5
6
7
+ – + – + – + – + –
INT
1AU/V
AUTO
ZERO
EVENT
LAMP
RECORDER
OUTPUT
REMOTE
100V
120V
220V
240V
1 2 3 4 5 6
OPEN
FILTER BYPASS
INT. OFFSET
REC. OFFSET
1V
100mV
10mV
RECORDER
F/S
8
9
10
11
12
13
14
15
17
16
Figure 4 Back Panel
Summary of Contents for ProStar 340
Page 2: ......
Page 42: ...16 03 914783 00 3 ...
Page 53: ...MAINTENANCE AND TROUBLESHOOTING ProStar 340 UV Vis Detector 27 Figure 10 Flowcell ...