38
03-914783-00:3
Flowcells:
Pathlengths from 0 mm to 10 mm, cell volumes from 0 mL
to 15 mL, stainless steel, titanium, or Kel-F, contact
materials, sapphire windows. 1000 psi pressure rating for
stainless steel cells, 500 psi pressure rating for Kel-F
cells, 2,000 psi pressure rating for variable pathlength
preparative cells; 7,000 psi pressure rating for high
pressure microbore cell.
Dimensions:
6
¼
inches high, 13
¼
inches deep, 9
¾
inches wide.
Weight
20 lbs.
Line Voltage
100, 120, 220, 240 Vac (± 10%), 50 or 60 Hz.
Parts and Accessories
P
ART
N
UMBER
D
ESCRIPTION
LAMPS
R000088515
Pre-aligned Deuterium Lamp Assembly
R000088516
Pre-aligned Tungsten Lamp Assembly
FLOWCELLS
R000088506
Conventional HPLC: 6 mm path, 9 µL volume, SS
R000088507
Microbore HPLC: 3 mm path, 1. µL volume, SS
R000088508
Semi-Prep/glass column: 3 mm path, 4.5 µL volume, SS
R000088509
Biocompatible HPLC: 6 mm, 9 µL, volume, Kel-F, rated to 500 psi
R000088510
Variable-path preparative: 0-3 mm path, 0-4.6 µL volume, SS, 1/8
inch fittings, rated to 2000 psi
R000088511
Conventional HPLC: 10mm path, 15 µL volume, SS
R000088512
Capillary HPLC and SFC: 240 µm path, 35 nL volume, borosilicate,
rated to 5000 psi
R000088513
Preparative biocompatible: pathlength adjusts from 0-3 mm, 0-4.6
µL volume, titanium, rated to 2000 psi
R000088514
High pressure microbore for packed cell SFC and LC/MS: 2 mm
path, 250 nL volume, SS, rated to 7000 psi
Summary of Contents for ProStar 340