7
Chapter 2
.
S
PECIFICATIONS
The specifications for the 4176 Programmable
µ
-Ohmmeter are listed in the following
paragraphs. In all cases the specifications are valid for full Kelvin Four-Terminal
measurements using connections having less than 20m
Ω
of lead resistance per wire.
2.1 Standard Measurement Mode Specifications
Table 2 – Range Characteristics
Range
Full Scale
Maximum
Resolution
Current
Source
1
20m
Ω
20.000m
Ω
1
µ
Ω
1A
.2
Ω
.30000
Ω
10
µ
Ω
1A
2
Ω
3.0000
Ω
100
µ
Ω
100mA
20
Ω
30.000
Ω
1m
Ω
10mA
200
Ω
300.00
Ω
10m
Ω
1mA
2k
Ω
3.0000k
Ω
100m
Ω
100
µ
A
20k
Ω
30.000k
Ω
1
Ω
10
µ
A
1
Current source is ±1%
absolute accuracy.
Table 3 – Standard Measurement Mode Accuracy
Accuracy
2
(± % of reading ± % of range)
Range
24 hours
1 Year
Temperature
Coefficient
3
20m
Ω
±0.006 ± 0.012
±0.02 ± 0.02
±0.002%/°C
.2
Ω
±0.006 ± 0.012
±0.02 ± 0.02
±0.002%/°C
2
Ω
±0.006 ± 0.012
±0.02 ± 0.02
±0.002%/°C
20
Ω
±0.006 ± 0.012
±0.02 ± 0.02
±0.002%/°C
200
Ω
±0.006 ± 0.012
±0.02 ± 0.02
±0.002%/°C
2k
Ω
±0.006 ± 0.012
±0.02 ± 0.02
±0.002%/°C
20k
Ω
±0.006 ± 0.012
±0.02 ± 0.02
±0.002%/°C
2
The accuracy
specifications listed are
valid following a 30 minute
warm-up at an ambient
temperature between 15
°
C
and 35
°
C, and include the
effects of line voltage
variations within the
allowed range.
3
Temperature coefficient
specified for temperature
ranges from 0°C to 15°C
and 35°C to 50°C.