8.4.1.2
Source Calibration
Step 1.
Short the SENSE HI and SENSE LO using a jumper.
Step 2.
Connect the digital voltmeter to the main board as follows:
DVM negative input to TP9
DVM positive input to TP10
Step 3.
Adjust RV1 for a DVM reading of -1.0000V.
Step 4.
Remove the DVM connection and the jumper.
Step 5.
Select the .2
Ω
range.
Step 6.
Connect the 4176 to a .1
Ω
standard resistor.
Step 7.
Adjust RV2 for a display reading equal to the value of the load.
Step 8.
Select the 2
Ω
range.
Step 9.
Connect the 4176 to a 1
Ω
standard resistor
Step 10.
Adjust RV3 for a display reading equal to the value of the load.
Step 11.
Select the 20
Ω
range.
Step 12.
Connect the 4176 to a 10
Ω
standard resistor
Step 13.
Adjust RV4 for a display reading equal to the value of the load.
Step 14.
Select the 200
Ω
range.
Step 15.
Connect the 4176 to a 100
Ω
standard resistor
Step 16.
Adjust RV5 for a display reading equal to the value of the load.
Step 17.
Select the 2k
Ω
range.
Step 18.
Connect the 4176 to a 1k
Ω
standard resistor
Step 19.
Adjust RV6 for a display reading equal to the value of the load.
Step 20.
Select the 20k
Ω
range.
Step 21.
Connect the 4176 to a 10k
Ω
standard resistor
Step 22.
Adjust RV7 for a display reading equal to the value of the load.
Step 23.
Reset the 4176 from the front panel
R
ESET
button or by turning off the power
switch and rebooting.
Step 24.
End of Souse Calibration.