- 57 -
HLCLO? QUERY
Reads the Hi-Lo Comparator low value from RAM for range.
Syntax:
HLCLO?
Response:
Floating (fixed) point Ohm value for current range.
"1.0000" for 1 Ohm in the 1 Ohm range and 100.00m for 100m Ohm in the
100m Ohm range, 1.0000k for 1k Ohm in the 1k Ohm range.
This command reads the value from RAM memory only. If you have used the
HLCLO command to write a HLC value it will be different than the value
stored in non-volatile memory if the SAVSETUP command has not been
sent.
Send RANGE command before sending this command.
Power-on default = Depends on range.
*IDN? QUERY
Returns the IEEE or RS-232 identification string from non-volatile.
Syntax:
*IDN?<crlf>
Response:
ID string
"VALHALLA SCIENTIFIC 4176,1.01G,0"
Example: *IDN?<crlf>
"VALHALLA SCIENTIFIC 4176,1.01G,0"<crlf>
ID_STRING : "VALHALLA SCIENTIFIC"
VERSION : "1.01G"
MODEL : "4176"
HARDWARE_VER : "0"
OPTION_STRING : "Option(s) : GPIB(IEEE488.2)"