7
2
When the instrument has completed the 10mV cal, the screen will prompt the following
message:
Step 5.
Apply 300mV using the voltage standard and press
E
NTER
to start the
range voltage calibration. Press the
C
LEAR
button if you wish to return to
the previous screen.
This step will perform the voltage calibration from 200m
Ω
range to the 20k
Ω
range.
When the instrument has completed the 300mV cal, the screen will prompt the following
message:
Step 6.
Press
E
NTER
to save the cal values or
C
LEAR
to return to the 300mV cal.
Step 7.
Press
E
NTER
to complete the calibration. The screen will display:
Sense calibration is now completed.