atttn<CR><LF>
a
Single-digit WMT700 address that corresponds to the first character of the
address
value
ttt
Period of time, in seconds, after which WMT700 has the measurement ready
n
Number of measurement values WMT700 calculates and returns as a response to one or
more subsequent
aD0!
commands
<CR><LF>
Terminates the response.
Example
1M! 10025<CR><LF>
Interpretation of the example message:
• Time after which the measurement is ready: 2 seconds
• Number of returned measurement values: 5
To fetch the measurement results, use the
aD0!
command.
5.26.1.8 aV! — Start Verification
This command starts verification. However, WMT700 does not return verification data
directly after this command. It sends a response that specifies when the verification results
can be retrieved using the
aD0!
command and the number of parameters. The WMT700
verification data includes a watchdog count, sensor measurement unit settings, and the
current SDI mode setting.
The command:
aV!
a
Single-digit WMT700 address that corresponds to the first character of the address
value
V
Starts verification command
!
Terminates the command
The response:
atttn<CR><LF>
Chapter 5 – Operation
159
Summary of Contents for WINDCAP WMT700 Series
Page 1: ...M211095EN H User Guide Vaisala WINDCAP Ultrasonic Wind Sensor Series WMT700 ...
Page 32: ...WMT700 User Guide M211095EN H 30 ...
Page 100: ...WMT700 User Guide M211095EN H 98 ...
Page 168: ...4 Remove the verifier WMT700 User Guide M211095EN H 166 ...
Page 174: ...WMT700 User Guide M211095EN H 172 ...
Page 182: ...WMT700 User Guide M211095EN H 180 ...
Page 194: ...WMT700 User Guide M211095EN H 192 ...
Page 198: ...WMT700 User Guide M211095EN H 196 ...
Page 201: ...Appendix G Certificate Appendix G Certificate 199 ...
Page 202: ...WMT700 User Guide M211095EN H 200 ...
Page 208: ...WMT700 User Guide M211095EN H 206 ...
Page 209: ......
Page 210: ...www vaisala com ...