7. Troubleshooting
7.1 Problem Situations
Problem
Probable Cause
Remedy
Connection to WMT700 is
lost.
Power supply is not sufficient.
Check that the power supply
matches the requirements listed in
Wind measurement failure.
WMT700 is sending irregular
data values.
The installation site is causing
measurement problems.
Check that the installation site
matches the requirements listed in
Snow, ice, a bird, or some other object is
blocking the measurement path
between the ultrasonic transducers.
Remove the blockage and check
that the wind transducers are not
damaged. If the blockage is ice or
snow, it melts after some time if
you are using a heated version of
WMT700. Duration of the blockage
depends on the severity of the
weather event. If birds are causing
the blockage, consider installing a
bird cage.
Reported wind direction is not
correct.
WMT700 is misaligned, which causes an
offset error.
Realign WMT700 according to
Wind direction indication acts
erratically.
Orientation of the sensor does not
match the
wndOrientation
parameter value.
Set the wndOrientation parameter
value according to the actual
orientation of the sensor. See
There is no response when
you try to switch to
configuration mode with the
OPEN
command.
The WMT700 settings are unknown or
they have been accidentally changed.
Reset the serial communication
settings to default values. See
Restoring Serial Port Settings
(page 170)
.
WMT700 does not respond to
any commands.
The communication profile is not
correct.
Configure the communication
profile to match the host
communication profile.
The wiring might be incorrect.
Connection works but data
messages are not available.
The command has been mistyped.
Use the
ERRORS
command to fetch
the error messages from
WMT700.See
Error Codes and Counts
(page 106)
.
Chapter 7 – Troubleshooting
167
Summary of Contents for WINDCAP WMT700 Series
Page 1: ...M211095EN H User Guide Vaisala WINDCAP Ultrasonic Wind Sensor Series WMT700 ...
Page 32: ...WMT700 User Guide M211095EN H 30 ...
Page 100: ...WMT700 User Guide M211095EN H 98 ...
Page 168: ...4 Remove the verifier WMT700 User Guide M211095EN H 166 ...
Page 174: ...WMT700 User Guide M211095EN H 172 ...
Page 182: ...WMT700 User Guide M211095EN H 180 ...
Page 194: ...WMT700 User Guide M211095EN H 192 ...
Page 198: ...WMT700 User Guide M211095EN H 196 ...
Page 201: ...Appendix G Certificate Appendix G Certificate 199 ...
Page 202: ...WMT700 User Guide M211095EN H 200 ...
Page 208: ...WMT700 User Guide M211095EN H 206 ...
Page 209: ......
Page 210: ...www vaisala com ...