For
condensable vapours
(water vapour, solvents, ...):
☞
Do not pump vapour before pump has reached its operat-
ing temperature and with gas ballast valve closed.
☞
Open gas ballast valve. Close gas ballast valve by turning
180°.
☞
The gas ballast valve is open if the flash on the gas ballast
cap shows away from the pump (MZ 2C VARIO / MD 4C
VARIO / PC 2002 VARIO), respectively downwards (PC 2003
VARIO / PC 2004 VARIO).
☞
With gas ballast valve open ultimate vacuum will be reduced,
pumping speed is decreased.
☞
Use inert gas at the air inlet to avoid the formation of explo-
sive mixtures.
In case of low boiling solvents when the formation of conden-
sate is unlikely, the use of gas ballast might be unnecessary.
☞
Operating the pump without gas ballast increases the sol-
vent recovery rates at the exhaust waste vapour condenser.
gas
ballast
Separation of condensate:
Both round bottom flasks are coated with a protective layer to
prevent disintegration in case of breakage or implosion.
☞
Collecting flask at the outlet: Remove joint clip, remove flask
and drain condensate.
☞
Separator at the inlet: Isolate or shut down the pump, admit
air or inert gas to atmospheric pressure. Remove joint clip,
remove flask and drain condensate.
Important:
Obey applicable regulations when disposing sol-
vents. Reuse if possible, purify if contaminated.
Check overpressure safety relief device at the exhaust waste
vapour condenser regularly, replace if necessary.
☞
Check especially for conglutination and cracks.
During operation PC 2002 VARIO / PC 2003 VARIO / PC 2004 VARIO:
overpressure
safety relief
device
catchpot at
the outlet
catchpot at
the inlet
Attention: Important notes regarding the use of gas ballast
☞
Make sure that air/gas inlet through the gas ballast valve never leads to hazard-
ous, explosive or otherwise dangerous mixtures. If in doubt, use inert gas.
☞
When using air rather than inert gas, risk of significant damage to equipment and/
or facilities, risk of personal injury or even loss of life exists due to the formation of
hazardous and/or explosive mixtures if air and pumped media react inside or at the
outlet of the pump.