page 75 of 92
If the pump does not achieve the ultimate vacuum:
- Whenever the diaphragms and valves have been replaced, a break-in
period of several hours is required before the pump achieves its ulti-
mate vacuum.
- In case of an unusual noise, switch off pump immediately and check
clamping disc positions.
If the specified ultimate vacuum is not achieved, and if this does not
change after the break-in period:
Check hose connectors at pump heads for leaks. If necessary recheck
valve seats and pump chambers.
Checking the ultimate vacuum
➨
After any intervention at the equipment (e.g., repair / maintenance)
the ultimate vacuum of the pump has to be checked. Only if the pump
achieves its specified ultimate vacuum, the pump’s leak rate is low
enough to ensure that no explosive atmospheres will occur in the inte-
rior of the equipment.
32
Summary of Contents for MZ 1C
Page 31: ...page 31 of 92 MZ 1C 3 2 7 10 11 6 9 8 ...
Page 32: ...page 32 of 92 MD 1C AK EK 10 3 6 1 8 2 9 7 15 16 17 13 11 8 14 19 18 ...
Page 33: ...page 33 of 92 MD 1C VARIO SP 7 10 6 12 11 8 Ensure sufficient venting of the pump ...
Page 64: ...page 64 of 92 4 3A 5 MD 1C size 2 2x 1 17 mm 14 mm max 1 4 ...
Page 66: ...page 66 of 92 6 7 8 4 6x Clean ...
Page 67: ...page 67 of 92 9 Replacing the diaphragm 10 11 20696828 20696838 ...
Page 68: ...page 68 of 92 12 1 2 3 13 14 Pay attention to number of washers ...
Page 69: ...page 69 of 92 15 16 17 Pay attention to number of washers ...
Page 70: ...page 70 of 92 18 1 2 19 20 4x Replacing the valves and assembling the pump heads ...
Page 71: ...page 71 of 92 21 22 23 ...
Page 73: ...page 73 of 92 27 28 1 2 28A size 2 2x 17 mm 10 mm MD 1C AK EK 5 1x MD 1C ...
Page 74: ...page 74 of 92 29 30 31 Assembling the connecting hose MD 1C AK EK 2x ...
Page 77: ...page 77 of 92 5 6 4 20638821 20638821 ...
Page 89: ...page 89 of 92 ...
Page 90: ...page 90 of 92 ...
Page 91: ...page 91 of 92 ...