Connect a gas-tight exhaust line at the pump
outlet if necessary. Always vent exhaust
gases appropriately (e.g., into a fume hood).
+
Never block the gas outlet. The exhaust
line must always be free of obstructions (no
back pressure) to ensure an unimpeded
discharge of gas. The cross-section of the
outlet tubing must be at least the size of the
pump’s exhaust connection.
✔
• Reduce the transmission of vibration. Prevent mechan
-
ical load due to rigid pipelines. Insert elastic hoses or
flexible elements as couplings between the pump and
rigid pipes.
Secure hose connections at the pump ap-
propriately, e.g., with hose clamps, to protect
against accidental detachment.
To reduce pump noise emanating from the
pump exhaust port, connect an exhaust hose
or use a silencer (see „Accessories“, pg. 56).
NOTICE
EX
✔
✘
NOTICE
Exhaust waste vapor condenser (EK) at the outlet
The
exhaust waste vapor condenser
enables an efficient condensation
of the pumped vapors at the outlet.
+
No backflow of condensates.
+
Controlled recovery of condensates.
+
Close to 100% solvent recovery.
+
The isolation cover protects against glass splinters in case of break-
age, acts as thermal isolation to avoid condensation of humidity and is
intended to absorb shocks.
Always install outlet tubing descending from
the pump or provide other measures to avoid
backflow of condensate towards the pump.
Summary of Contents for MZ 1C
Page 31: ...page 31 of 92 MZ 1C 3 2 7 10 11 6 9 8 ...
Page 32: ...page 32 of 92 MD 1C AK EK 10 3 6 1 8 2 9 7 15 16 17 13 11 8 14 19 18 ...
Page 33: ...page 33 of 92 MD 1C VARIO SP 7 10 6 12 11 8 Ensure sufficient venting of the pump ...
Page 64: ...page 64 of 92 4 3A 5 MD 1C size 2 2x 1 17 mm 14 mm max 1 4 ...
Page 66: ...page 66 of 92 6 7 8 4 6x Clean ...
Page 67: ...page 67 of 92 9 Replacing the diaphragm 10 11 20696828 20696838 ...
Page 68: ...page 68 of 92 12 1 2 3 13 14 Pay attention to number of washers ...
Page 69: ...page 69 of 92 15 16 17 Pay attention to number of washers ...
Page 70: ...page 70 of 92 18 1 2 19 20 4x Replacing the valves and assembling the pump heads ...
Page 71: ...page 71 of 92 21 22 23 ...
Page 73: ...page 73 of 92 27 28 1 2 28A size 2 2x 17 mm 10 mm MD 1C AK EK 5 1x MD 1C ...
Page 74: ...page 74 of 92 29 30 31 Assembling the connecting hose MD 1C AK EK 2x ...
Page 77: ...page 77 of 92 5 6 4 20638821 20638821 ...
Page 89: ...page 89 of 92 ...
Page 90: ...page 90 of 92 ...
Page 91: ...page 91 of 92 ...